Measurements of the relative output factors for CyberKnife collimators.

نویسندگان

  • Cheng Yu
  • Gabor Jozsef
  • Michael L J Apuzzo
  • Zbigniew Petrovich
چکیده

OBJECTIVE To determine accurately the relative output factors, defined as the ratio of the nominal dose rate for a given collimator to that of the 60-mm collimator. This is particularly important for radiosurgical treatment of functional disorders, such as trigeminal neuralgia, in which a single large radiation dose is delivered to the target with a small collimator, such as the 5-mm collimator for CyberKnife radiosurgery. Numerous studies on the output factors have been reported for the Leksell gamma knife unit but none for the CyberKnife system. METHODS Measurements of the relative output factors for all 12 collimators were performed by three different methods: silicon diode, radiographic film, and thermoluminescent dosimetry microcubes. The silicon diode is designed for measurements in small (1-50 MV) photon beams performed in water or air. Film and thermoluminescent dosimetry measurements were performed in a plastic phantom. RESULTS The measured relative output factors for the three methods were very similar except for the three smallest collimators (5, 7.5, and 10 mm). The measured difference between the above methods was approximately 2%. The mean value of the output factor for the 5-mm collimator was 0.686 +/- 0.024. The uncertainties of the output factors are expected to increase with the decrease of collimator diameter. They range from approximately 1 to 4% of the relative output factor. CONCLUSION The relative output factor can be measured with an acceptable accuracy even for the smallest (5-mm) CyberKnife collimators. This requires the selection of appropriate dosimetric detectors and measuring procedures. The results obtained with the diode are considered more accurate than with the other two methods.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of basic physical and dosimetric parameters of synthetic single-crystal diamond detector and its use in Leksell Gamma Knife and CyberKnife small radiosurgical fields

Background: To determine the basic physical and dosimetric properties of a new synthetic single-crystal diamond detector and its application for relative small field dosimetry. Materials and Methods: The pre-irradiation dose required to stabilize detector response, dose rate dependence, photon and electron energy dependence, temperature dependence and angular dependence of MicroDiamond detector...

متن کامل

Use of novel fibre-coupled radioluminescence and RADPOS dosimetry systems for total scatter factor measurements in small fields.

A dosimetry system based on Al2O3:C radioluminescence (RL), and RADPOS, a novel 4D dosimetry system using microMOSFETs, were used to measure total scatter factors, (S(c,p))(f(clin))(det), for the CyberKnife robotic radiosugery system. New Monte Carlo calculated correction factors are presented and applied for the RL detector response for the 5, 7.5 and 10 mm collimators in order to correct for ...

متن کامل

Total scatter factors of small beams: a multidetector and Monte Carlo study.

The scope of this study was to estimate total scatter factors (S(c,p)) of the three smallest collimators of the Cyberknife radiosurgery system (5-10 mm in diameter), combining experimental measurements and Monte Carlo simulation. Two microchambers, a diode, and a diamond detector were used to collect experimental data. The treatment head and the detectors were simulated by means of a Monte Carl...

متن کامل

Monte Carlo simulated corrections for beam commissioning measurements with circular and MLC shaped fields on the CyberKnife M6 System: a study including diode, microchamber, point scintillator, and synthetic microdiamond detectors.

Monte Carlo simulation was used to calculate correction factors for output factor (OF), percentage depth-dose (PDD), and off-axis ratio (OAR) measurements with the CyberKnife M6 System. These include the first such data for the InCise MLC. Simulated detectors include diodes, air-filled microchambers, a synthetic microdiamond detector, and point scintillator. Individual perturbation factors were...

متن کامل

Small field measurements with a novel silicon position sensitive diode array.

DOSI, a novel dosimeter based on position sensitive detectors for particle physics experiments, was used for relative clinical dosimetry measurements in small radiotherapy fields. The device is capable of dynamic measurements in real time and provides sub-millimetre spatial resolution. The basic beam data for a stereotactic radiotherapy collimator system (BrainLAB) using 6 MV photons were measu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Neurosurgery

دوره 54 1  شماره 

صفحات  -

تاریخ انتشار 2004