Quantification of the specific membrane capacitance of single cells using a microfluidic device and impedance spectroscopy measurement.
نویسندگان
چکیده
The specific membrane capacitance (SMC) is an electrical parameter that correlates with both the electrical activity and morphology of the plasma membrane, which are physiological markers for cellular phenotype and health. We have developed a microfluidic device that enables impedance spectroscopy measurements of the SMC of single biological cells. Impedance spectra induced by single cells aspirated into the device are captured over a moderate frequency range (5 kHz-1 MHz). Maximum impedance sensitivity is achieved using a tapered microfluidic channel, which effectively routes electric fields across the cell membranes. The SMC is extracted by curve-fitting impedance spectra to an equivalent circuit model. From our measurement, acute myeloid leukemia (AML) cells are found to exhibit larger SMC values in hypertonic solutions as compared with those in isotonic solutions. In addition, AML cell phenotypes (AML2 and NB4) exhibiting varying metastatic potential yield distinct SMC values (AML2: 16.9 ± 1.9 mF/m(2) (n = 23); NB4: 22.5 ± 4.7 mF/m(2) (n = 23)). Three-dimensional finite element simulations of the microfluidic device confirm the feasibility of this approach.
منابع مشابه
Development of a Microfluidic Device for Single Cell Specific Membrane Capacitance Quantification
The specific membrane capacitance (SMC) of biological cell membranes correlates with cells’ electrical activity and morphology, which are physiological markers for cellular phenotype and health. Conventionally, SMC measurements are conducted using electro-rotation and Patchclamping, which entail long time training and stringent operation skills. Both techniques also suffer from limited throughp...
متن کاملFast curve fitting using neural networks
Related Articles Optimization of the configuration of pixilated detectors based on the Shannon-Nyquist theory Rev. Sci. Instrum. 83, 10E139 (2012) Multiscale reactive molecular dynamics J. Chem. Phys. 137, 22A525 (2012) Quantification of the specific membrane capacitance of single cells using a microfluidic device and impedance spectroscopy measurement Biomicrofluidics 6, 034112 (2012) Geometri...
متن کاملEquivalent Electrical Circuit Modeling of Ceramic-Based Microbial Fuel Cells Using the Electrochemical Impedance Spectroscopy (EIS) Analysis
The effect of the thickness of ceramic membrane on the productivity of microbial fuel cells (MFCs) was investigated with respect to the electricity generation and domestic wastewater treatment efficiencies. The thickest ceramic membrane (9 mm) gained the highest coulombic efficiency (27.58±4.2 %), voltage (681.15±33.1 mV), and current and power densities (447.11±21.37 mA/m2, 63.82±10.42 mW/m2) ...
متن کاملThe Instrumentation of a Microfluidic Analyzer Enabling the Characterization of the Specific Membrane Capacitance, Cytoplasm Conductivity, and Instantaneous Young’s Modulus of Single Cells
This paper presents the instrumentation of a microfluidic analyzer enabling the characterization of single-cell biophysical properties, which includes seven key components: a microfluidic module, a pressure module, an imaging module, an impedance module, two LabVIEW platforms for instrument operation and raw data processing, respectively, and a Python code for data translation. Under the contro...
متن کاملDynamic monitoring of single cell lysis in an impedance-based microfluidic device
A microfluidic device that is capable of trapping and sensing dynamic variations in the electrical properties of individual cells is demonstrated. The device is applied to the real-time recording of impedance measurements of mouse embryonic stem cells (mESCs) during the process of membrane lysis, with the resulting changes in the electrical properties of cells during this process being quantita...