Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An In Vitro Study

نویسندگان

  • Masoud Saatchi
  • Shiva Iravani
  • Mehrdad Akhavan Khaleghi
  • Amin Mortaheb
چکیده

INTRODUCTION The aim of this in vitro study was to evaluate the correlation between accuracy of Root ZX electronic foramen locator and root canal curvature. METHODS AND MATERIALS One hundred and ten extracted mandibular molars were selected. Access cavity was prepared and coronal enlargement of mesiobuccal canal was performed. A #10 Flexofile was inserted into the mesiobuccal canal, and a radiography was taken to measure the degree of curvature by Schneider's method. The actual working length (AWL) was defined by inserting the file until its tip could be observed at a place tangential to the major apical foramen and then 0.5 mm was subtracted from this measurement. For the electronic working length (EWL) measurement, the apical 3 or 4 mm of the root was embedded in alginate as the electrolyte material. The file was inserted into the root canal to the major foramen, until the APEX reading was shown on the electronic device and then pulled back until the visual display showed the 0.5-mm mark. The AWL was subtracted from the EWL to define the distance between the file tip and the point 0.5 mm coronal to the major apical foramen. Data were analyzed using the Pearson's correlation coefficient. RESULTS The accuracy of Root ZX within ±0.1 mm and ±0.5 mm was 38.2% and 94.6%, respectively. There was no correlation between the distance from the EWL to the AWL and the degree of root canal curvature (r=0.097, P=0.317). CONCLUSION Root canal curvature did not influence the accuracy of Root ZX foramen locator.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

ارزیابی دقت یک نوع Electronic Apex Locator در تعیین طول کانال دندانهای شیری در شرایط In vivo

Background and Aim: Radiography is the most commonly used technique in root canal length determination, but its application in pedodontics is difficult due to many problems such as radiation hazards, superimposition of permanent teeth buds and uncooperative children. The aim of this study was to investigate the accuracy of an electronic apex locator (Dentaport ZX) in measuring the root canal le...

متن کامل

Comparison of Working Length Determination Using Apex Locator and Conventional Radiography in Curved Canals-an In Vitro Study

Introduction: The exact measurement of the root canal length is of great importance in root canal therapy. Although, determination of the root canal length through radiography is the most common method, recently electronic apex locators are being used to determine working length and decrease the number of radiographs. The purpose of this study was to compare the accuracy of an electronic apex l...

متن کامل

In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination

In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...

متن کامل

The effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants

BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...

متن کامل

ارزیابی دقت یک نوع Electronic Apex Locator  در تعیین طول دندانهای شیری (In-vitro)

Statement of Problem: Radiography is the most common technique in working length determination, however, because of its limitations is not considered as an ideal technique. Its application, particularly for children due to radiation hazards, technical problems in young and unco-operative children and the superimposition of permanent teeth bud on primary teeth root, lead to numerous problems.Pur...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2017