Oversensing of atrial fibrillatory waves in a subcutaneous implantable cardioverter-defibrillator

نویسندگان

  • Ankur A. Karnik
  • Robert H. Helm
  • Kevin M. Monahan
چکیده

Introduction Implantable cardioverter-defibrillators (ICDs) have been shown to improve survival in patients with ischemic or nonischemic cardiomyopathy when used for primary and secondary prevention settings. The limitations of transvenous ICDs include postimplant adverse events, defibrillator lead failure, and inappropriate shocks for atrial tachyarrhythmias due to fast ventricular response rate that are associated with an increase in all-cause mortality. Given these shortcomings, subcutaneous ICDs (S-ICDs) may offer several potential advantages. Prior S-ICD studies have shown a complication rate and shock efficacy comparable to transvenous systems and significantly better supraventricular tachycardia (SVT) discrimination compared to transvenous ICDs. Inappropriate shocks in S-ICDs are almost exclusively due to double counting secondary to Twave oversensing. We report a patient with an S-ICD and hypertrophic cardiomyopathy (HCM) who received an inappropriate shock for atrial fibrillation due to oversensing of atrial fibrillatory waves. This has not been reported previously.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Atrial Sensing and Blanking in Dual-Chamber Implantable Cardioverter-Defibrillators

Atrial Sensing and Blanking in Dual-Chamber Implantable Cardioverter-Defibrillators Accurate atrial sensing during tachycardia is essential for reliable determination of atrial rate, which is a critical component of interval-based dual-chamber supraventricular tachycardia (SVT)–VT discrimination algorithms. Thus, atrial sensing in dual-chamber implantable cardioverter-defibrillators (ICDs) must...

متن کامل

Routine exercise testing could not predict T‐wave oversensing in a patient after a subcutaneous implantable cardioverter‐defibrillator implant

Subcutaneous implantable cardioverter-defibrillators (S-ICDs) are susceptible to T-wave oversensing (TWOS) caused by high rate-dependent QRS-T morphology changes. We experienced an inappropriate S-ICD shock due to TWOS, which could not be predicted by routine exercise testing. A newly available high-pass filter might be effective for avoiding this.

متن کامل

Subcutaneous Implantable Cardioverter Defibrillator: Early Experience

INTRODUCTION The implantable cardioverter defibrillator had been increasing the survival of patients at high risk for sudden cardiac death. The subcutaneous implantable cardioverter defibrillator was developed to mitigate the complications inherent to lead placement into cardiovascular system. OBJECTIVE To report the initial experience of 18 consecutive cases of subcutaneous implantable cardi...

متن کامل

Changing place, changing future: Repositioning a subcutaneous implantable cardioverter-defibrillator can resolve inappropriate shocks secondary to myopotential oversensing

Introduction The clinical benefits of implantable cardioverter-defibrillator (ICD) therapy have been widely documented in clinical studies of selected patient populations. Subcutaneous ICD (S-ICD) is a valid alternative to transvenous ICDinpatientswhodonot require cardiac resynchronization or antibradycardia or antitachycardia pacing, and has the additional advantage of avoiding possible acute ...

متن کامل

Ventricular tachycardia slower than the rate cut-off of a subcutaneous cardiac defibrillator sensed and successfully treated as a result of oversensing

Introduction The subcutaneous implantable cardioverter-defibrillator system (S-ICD) has been an important alternative to transvenous defibrillators for certain clinical situations, such as when there is a venous access problem or high risk of bacterial endocarditis. This device reliably detects ventricular tachyarrhythmias and effectively delivers shocks. However, oversensing is more common wit...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 3  شماره 

صفحات  -

تاریخ انتشار 2017