Adaptation of the troponin T ELISA test to a microplate immunoassay reader.

نویسندگان

  • I Penttilä
  • K Hirvonen
  • A Julkunen
  • K Penttilä
  • T Rantanen
چکیده

Troponin T concentration in serum is usually measured by the automated method developed by Boehringer Mannheim for the ES-series of analysers. These instruments need at least 140 microliters of serum and 700 microliters of reagents for a single analysis, which takes 90 min. We describe an alternative procedure, using streptavidin-coated microtitre plates, troponin T reagents of Boehringer Mannheim and an ELISA-reader to measure the concentration of troponin T. The present assay needs only 30 microliters of sample and 200 microliters of reagents, and it takes 75 min; the detection limit is 0.10 microgram/l. We also assessed the microtitre plate method for sensitivity and precision and compared the results with those measured by an ES-300 automatic analyser. Both methods have the same measurement range for troponin T of 0.1 to 15 micrograms/l. For daily routine use of the microtitre plate method we recommend duplicate determinations.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the troponin T and troponin I ELISA tests, as measured by microplate immunoassay techniques, in diagnosing acute myocardial infarction.

We describe an improved procedure using a standard microplate immunoassay reader to measure the concentration of troponin T in human serum. We also describe an immunoassay for troponin I in serum. Only 160 microliters of serum are needed for a single analysis of each troponin. For comparison, creatine kinase MB mass analysis in serum was performed with a commercial luminometric method. From 95 ...

متن کامل

Cavity-Enhanced Immunoassay Measurements in Microtiter Plates Using BBCEAS.

We report on the first detailed use of broadband cavity enhanced absorption spectroscopy (BBCEAS) as a detection system for immunoassay. A vertical R ≥ 0.99 optical cavity was integrated with a motorized XY stage, which functioned as a receptacle for 96-well microtiter plates. The custom-built cavity enhanced microplate reader was used to make measurements on a commercially available osteocalci...

متن کامل

Collection and analysis of complex ELISA data using a microcomputer.

A system for automatic Enzyme-Linked Immunosorbent Assay (ELISA) data collection and processing is described. Hardware for the system consists of a Dynatech microplate reader and Apple II series microcomputer. Software developed by the authors allows automatic configuration of the reader and direct data input from the microplate reader to the microcomputer. After optical density data is input t...

متن کامل

Application of Elisa Microplate Reader in Complement Fixation Test Used for Dourine Diagnosis. a Preliminary Study

The aim of the present work was to estimate the possibility of improving the objectiveness of complement fixation test (CFT) in the diagnosis of dourine by using an ELISA microplate reader. The investigations were carried out on control positive and negative sera and 16 sera obtained for a reference examination. Six modifications of CFT performing and reading methods were compared (tube, microp...

متن کامل

Quantitative bedside assay for cardiac troponin T: a complementary method to centralized laboratory testing.

BACKGROUND In the evaluation of chest pain patients, whole blood bedside assays of highly specific cardiac molecules may be an attractive alternative to centralized clinical chemistry testing. We now report on an optimized test strip device for cardiac troponin T (cTnT) that can be analyzed by a cardiac reader for quantitative assessment of the test result. METHODS AND RESULTS The cTnT test s...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies

دوره 33 1  شماره 

صفحات  -

تاریخ انتشار 1995