A novel method for measuring the erythrocyte deformability, under oscillatory shear stress.
نویسندگان
چکیده
A novel apparatus for detecting a subtle decrease in erythrocyte deformability was developed. The oscillatory shape change of erythrocytes was monitored under an oscillatory shear stress (modulating the oscillation frequency in the range of 9-90 dyn/cm2). The ellipsoidal deformation of intact erythrocytes perfectly followed the oscillatory force of up to about 2 Hz, but the diamide-treated cells showed a phase difference of the oscillatory deformation at 1.7 Hz.
منابع مشابه
Interpretation of osmotic gradient ektacytometry (osmoscan) data: a comparative study for methodological standards.
Osmotic gradient ektacytometry (measuring elongation index in the function of osmolality at a constant shear stress) is a sensitive method to analyze red blood cell (RBC) deformability and investigating the optimal osmolality range for the cells in normal or pathophysiological cellular and micro-environmental conditions. However, the methodological conditions are different, since the results ar...
متن کاملRed blood cell deformability and aggregation behaviour in different animal species.
Comparative animal studies showed the wide variation of whole blood and plasma viscosity, and erythrocyte aggregation among mammalian species. Whole blood viscosity and red blood cell aggregation is influenced by red cell fluidity. To evaluate differences in erythrocyte deformability in mammals, three species were investigated, whose erythrocytes have a different aggregation property: horse, as...
متن کاملMeasuring Deformability and Red Cell Heterogeneity in Blood by Ektacytometry.
Decreased red cell deformability is characteristic of several disorders. In some cases, the extent of defective deformability can predict severity of disease or occurrence of serious complications. Ektacytometry uses laser diffraction viscometry to measure the deformability of red blood cells subject to either increasing shear stress or an osmotic gradient at a constant value of applied shear s...
متن کاملEffect of Aging on Deformability of Erythrocytes in Shear Flow
To study the effect of aging on deformability of an erythrocyte, rheological measurement has been performed after exposure to a shear field in vitro. Deformability was evaluated with shear stress responsiveness and with critical elongation calculated from an exponential curve between an elongation index and shear stress. Human erythrocytes were classified according to their density by a centrif...
متن کاملBlood storage alters mechanical stress responses of erythrocytes
BACKGROUND Erythrocytes undergo irreversible morphological and biochemical changes during storage. Reduced levels of deformability have been reported for stored erythrocytes. Erythrocyte deformability is essential for healthy microcirculation. OBJECTIVE The aim of this study is to evaluate shear stress (SS) induced improvements of erythrocyte deformability in stored blood. METHODS Deformabi...
متن کاملذخیره در منابع من