Transcutaneous PCO2 Measurement at Low Temperature for Reliable and Continuous Free Flap Monitoring: Experimental and Clinical Study
نویسندگان
چکیده
BACKGROUND Measurement of transcutaneous oxygen pressure (TcPO2) and transcutaneous carbon dioxide pressure (TcPCO2) has been used for free flap monitoring. Because these values are obtained with sensor probes heated to 44°C, there is potential for low-temperature burns on skin flaps. We measured TcPO2 and TcPCO2 at 37°C in both animals and humans to determine the feasibility and safety of the procedure as a postoperative flap monitoring method. METHODS Twelve epigastric island flaps were elevated in rabbits, and TcPO2 and TcPCO2 were measured at 37°C before and after ligation of the pedicles. In addition, TcPO2 and TcPCO2 at 37°C were measured in healthy men. Subsequently, the method was applied to postoperative monitoring of free flaps in 49 clinical cases. RESULTS TcPO2 and TcPCO2 values were significantly affected by the experimental flap elevation. A rapid increase in TcPCO2 was observed with both arterial and venous occlusion. In the healthy men, TcPO2 and TcPCO2 were measurable at all skin surface sites. In the clinical cases of free flap transfer, TcPO2 values remained very low for at least 72 hours. TcPCO2 values ranged from 40 to 70 mm Hg for 72 hours in more than 80% of cases. In 2 cases, TcPCO2 values increased to more than 90 mm Hg, and exploration surgery was performed. These compromised flaps were saved by reanastomosis of the veins. CONCLUSIONS Continuous monitoring of TcPCO2 at 37°C can provide objective information and alert doctors and nurses to the need for checking the free flap.
منابع مشابه
Comparison of Transcutaneous and Capillary Measurement of PCO2 in Hypercapnic Subjects.
BACKGROUND Measurement of PCO2 is vital in determining effective alveolar ventilation. However, obtaining capillary PCO2 by a skin prick of the earlobe is painful, and nocturnal measurements disturb sleep. End-expiratory measurement of PCO2 is also well established, but there is a low precision in predicting arterial or capillary CO2. The purpose of the study was to evaluate nocturnal measureme...
متن کاملComparison of ventilator-integrated end-tidal CO2 and transcutaneous CO2 monitoring in home-ventilated neuromuscular patients.
BACKGROUND Non-invasive transcutaneous capnometry (TcCO2) is used to assess the home ventilation's efficiency. Recently, end-tidal CO2 (ETCO2) sensors have been integrated in life-support home ventilators. The purpose of this study was to compare the ventilator-integrated ETCO2 with TcCO2, in home-ventilated neuromuscular disease patients. METHODS ETCO2 and TcCO2 were simultaneously measured ...
متن کاملAccuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
Background. In premature infants, maintaining blood partial pressure of carbon dioxide (pCO2) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO2), compared to blood partial pressure of carbon dioxide (pCO2). Methods. Retrospective observational study in a tertiary neonatal intensive ...
متن کاملTranscutaneous and capillary pCO2 and pO2 measurements in healthy adults.
Transcutaneous carbon dioxide and oxygen tensions (tc-pCO2 and tc-pO2) were measured in seven healthy adult volunteers during hyperventilation in atmospheric air and during CO2 inhalation. Three skin sensors were applied to each subject: an O2 electrode, a CO2 electrode, and a combined O2-CO2 electrode, each heated to 44 degrees C. We observed close correlation between tc-pCO2 and capillary-pCO...
متن کاملWhat is the potential role of transcutaneous carbon dioxide in guiding acute noninvasive ventilation?
BACKGROUND Transcutaneous carbon dioxide (PtcCO2 ) monitoring is rarely used in the acute hospital setting, where serial samples of arterial blood are instead taken to measure carbon dioxide tension (PaCO2 ). In this pilot observational study, we assessed the potential of PtcCO2 monitoring to calculate pH and guide management of acute noninvasive ventilation (NIV). METHODS Ten subjects with a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 1 شماره
صفحات -