A High-pass Birdcage Coil for Small Animal Imaging at 600MHz/14.1T

نویسندگان

  • Arthur W. Magill
  • Hongxia Lei
  • Rolf Gruetter
چکیده

Introduction: Birdcage coils become difficult to build at high field strength, as smaller capacitor values are required and coil sections become large relative to the RF wavelength. This abstract demonstrates a birdcage coil operating at 600MHz (λfree space=0.5m) for small animal imaging. Methods: A 12-rung high-pass design was chosen (50mm diameter × 32mm long, 2mm track width; 120mm diameter × 200mm long shield). Probe dimensions were chosen to allow construction from off-the-shelf plastic tubes (Angst & Pfister). Birdcage Builder was used to calculate initial capacitor values [1]. The probe was then simulated using Microwave Studio (CST) to fine-tune the capacitor values and investigate B1 field homogeneity. The probe length was adjusted slightly to move the frequency of the end-ring mode away from that of the desired birdcage mode. The conducting structure was manufactured from flexible PCB (0.1mm FR4, 35μm Cu). Five variable capacitors were used (2×tuning, 2×matching, balance, 1-9pF, type NMKJ10HVE, Voltronics); fixed capacitors are 8.2pF, type 100B (ATC). Bazooka baluns were fitted to the coaxial cables approximately 40mm from the probe, (9mm diameter × 50mm long PTFE tubes coated with 35μm copper tape (3M), tuned using three 1.2pF capacitors (100B, ATC)). The coil was characterised using a network analyser (E5071C, Agilent). MR data was acquired using a 14.1T/26cm horizontal bore scanner (Agilent). B1 maps were generated using the double-angle method (60°/120° FA, 0.2mm slice thickness, TR/TE =20,000/2.5ms, axial FOV: 40×40mm, 128×128 matrix, coronal FOV: 40×80mm, 128×256 matrix) [2]. Images of an ex-vivo mouse head were acquired using a fast spin-echo sequence (TR/TE=6000/10ms, 256×256 matrix, 20×20mm FOV, 15 × 0.1mm slices). Results: Good correspondence was found between simulated and measured probe reflection (fig. 1). The loaded birdcage was matched to -42.6dB and -39.7dB; coupling between channels was -19.5dB. Unloaded and loaded Q-factors (table 1) demonstrate good sample loading. Figs. 2 and 3 show simulated and measured B1 maps, demonstrating good axial homogeneity, reasonable coverage along the z-axis, and a slight central brightening effect. Discussion: A high-pass birdcage coil has been shown to produce a homogeneous B1 field over a field-of-view large enough for mouse and small rat imaging at 600MHz, despite the short RF wavelength. The coil was designed as a transmit probe; image SNR is sub-optimal, as preamplifier is located outside the magnet bore, introducing ca. 2dB cable losses before the preamplifier. Future work will focus on developing preamplifiers for use inside the magnet bore, and adding active detuning to the coil to allow use of surface and array receive coils. References: [1] C. L. Chin et al, Birdcage Builder, CNR, Penn State 1998; [2] Insko et al, JMR A 103: 82–85 (1993); Acknowledgements: This study was supported by the Centre d’Imagerie BioMédicale (CIBM) of the UNIL, UNIGE, HUG, CHUV, EPFL and the Leenaards and Jeantet Foundations.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Quadrature Lowpass Birdcage Coil for Low Field Open Mri Scanner

The main goal of this research was to demonstrate the utility of using a quadrature birdcage coil in lowpass version for magnetic resonance imaging (MRI) in open scanner at very low static field. When used in horizontal bore MRI systems, the birdcage coil is able to produce circular polarized field using quadrature excitation, thus reaching SNR improvement of 2 as maximum value. The birdcage co...

متن کامل

Design of Crisscrossed Double-Layer Birdcage Coil for Improving B1 + Field Homogeneity for Small-Animal Magnetic Resonance Imaging at 300 MHz

We design a crisscrossed double-layer birdcage (DLBC) coil by modifying the coil geometry of a standard single-layer BC (SLBC) coil to enhance the homogeneity of transmitting magnetic flux density (B1 ) along the main magnetic field (B0)-direction for small-animal magnetic resonance imaging (MRI) at 300 MHz. The performance assessment of the crisscrossed DLBC coil is conducted by computational ...

متن کامل

An Asymmetric Birdcage Coil for Small-animal MR Imaging at 7T

The birdcage (BC) coil is currently being utilized for uniform radiofrequency (RF) transmit/receive (Tx/Rx) or Tx-only configuration in many magnetic resonance (MR) imaging applications, but insufficient magnetic flux (|B1|) density and their non-uniform distribution still exists in high-field (HF) environments. We demonstrate that the asymmetric birdcage (ABC) transmit/receive (Tx/Rx) volume c...

متن کامل

Magnetron volume coil for Magnetic Resonance Imaging of rodents

Introduction. There is still a great interest to develop volume coils for small-animal MRI. A transceiver volume coil with slotted end-rings and similar to the standard birdcage is proposed in this work. This coil design resembles Mansfield ́s cavity resonator [1]. Unlike the traditional birdcage coil, our coil design was formed with two slotted surface coils [2] connected via rungs like the tra...

متن کامل

BirdcageBuilder: Design of Specified-Geometry Birdcage Coils with Desired Current Pattern and Resonant Frequency.

A simple method for birdcage coil design (high pass, low pass, and band pass) is presented. Rather than iteratively approaching the Larmor resonant frequency with known capacitances and calculated inductances, a more versatile approach to birdcage coil design is developed and validated, in which the necessary capacitances are calculated using a desired current pattern, a calculated inductance, ...

متن کامل