Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients.

نویسندگان

  • Dennis S C Lam
  • Dexter Y L Leung
  • Thomas Y H Chiu
  • Dorothy S P Fan
  • Eva Y Y Cheung
  • Tien-Yin Wong
  • Jimmy S M Lai
  • Clement C Y Tham
چکیده

PURPOSE To evaluate whether the pressure phosphene tonometer (PPT) is suitable for self-tonometry in patients with glaucoma or ocular hypertension. METHODS This was a prospective comparative study of 102 eyes of 102 patients with chronic glaucoma or ocular hypertension. Intraocular pressure (IOP) measurements by the Goldmann tonometer (GT) were compared with self-measured readings with the PPT. Patients evaluated the ease of home use of the PPT. The last 15 patients were asked to stop their glaucoma medications, and the ability of the PPT to detect an elevated IOP during self-tonometry was studied. RESULTS The mean +/- SD difference between PPT and GT readings was -0.24 +/- 1.57 mm Hg. Comparing the PPT with the GT, 86% of the readings were within +/- 2.0 mm Hg, and 91% were within +/- 3.0 mm Hg. Spearman's correlation coefficient was 0.91. A Bland-Altman plot showed that the 95% limits of agreement between the two methods lay between 2.90 and -3.38 mm Hg. Within-subject coefficients of variation for the GT and the PPT were 4.4% and 7.3%, respectively. In detecting an elevated IOP of more than 21 mm Hg, the sensitivity and the specificity of the PPT were 92.3% and 98.6% respectively. The mean satisfaction score for home use of the PPT was 87.4 +/- 16.3 (maximum 100). CONCLUSIONS With proper training and technique, self-tonometry with the PPT appears to be accurate up to at least 25 mm Hg and is reproducible. The PPT was sensitive and specific in detecting an elevated IOP of more than 21 mm Hg. As patients were expected to seek ophthalmic care before the self-measured IOP reaches 25 mm Hg, the PPT may have a value for self-monitoring. Patients rated the PPT as satisfactory for home use. Because the PPT is portable and relatively inexpensive and requires no topical anesthesia or direct contact with the eyeball, it may have potential as an instrument for home self-tonometry.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clinical evaluation of the pressure phosphene tonometer in patients with glaucoma.

AIM To evaluate the reliability of the pressure phosphene tonometer in comparison with the Goldmann applanation tonometer. METHODS 45 consecutive patients with glaucoma (78 eyes) participated in the study. Eyes with previous eye surgery, a documented peripheral visual field defect, a refractive error of more than 5 dioptres, and patients who were unable to understand the procedure were exclud...

متن کامل

Pressure phosphene tonometry versus Goldmann applanation tonometry for measuring intraocular pressure before and after LASIK.

PURPOSE To compare pressure phosphene tonometry with Goldmann applanation tonometry for measuring intraocular pressure (IOP) before and after LASIK. METHODS Forty-three (18 men and 25 women) consecutive healthy patients underwent complete pre- and postoperative LASIK ophthalmologic assessments including manifest and cycloplegic refraction, keratometry, and central corneal thickness. Three rep...

متن کامل

مقایسه دقت اندازه گیری فشار داخل چشم با سه تونومتر پالس ایر، تونوپن و گلدمن

Introduction & Objective: The measurement of Intraocular pressure (IOP) is very important not only in the diagnosis and management of glaucomatous conditions but also in any ophthalmologic examination. So this study was designed to compare the results of noncontact pulsair and tononpen tonometry with goldmann known as gold standard method. Materials & Methods: In this cross sectional study o...

متن کامل

Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients includ...

متن کامل

Applanation tonometry: a comparison of the Perkins handheld and Goldmann slit lamp-mounted methods

PURPOSE To compare intraocular pressure (IOP) measurements, taken using Perkins applanation tonometry (PAT) and Goldmann applanation tonometry (GAT). METHODS 100 eyes of 100 patients underwent Perkins and Goldmann applanation tonometry, with a randomized order of modality, performed by a masked observer. The right eye was measured, for all subjects, and the data used in statistical analysis. ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Investigative ophthalmology & visual science

دوره 45 9  شماره 

صفحات  -

تاریخ انتشار 2004