Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning.
نویسندگان
چکیده
BACKGROUND AND PURPOSE To investigate the accuracy of photon dose calculations performed by the Anisotropic Analytical Algorithm, in homogeneous and inhomogeneous media and in simulated treatment plans. MATERIALS AND METHODS Predicted dose distributions were compared with ionisation chamber and film measurements for a series of increasingly complex situations. Initially, simple and complex fields in a homogeneous medium were studied. The effect of inhomogeneities was investigated using a range of phantoms constructed of water, bone and lung substitute materials. Simulated treatment plans were then produced using a semi-anthropomorphic phantom and the delivered doses compared to the doses predicted by the Anisotropic Analytical Algorithm. RESULTS In a homogeneous medium, agreement was found to be within 2% dose or 2mm dta in most instances. In the presence of heterogeneities, agreement was generally to within 2.5%. The simulated treatment plan measurements agreed to within 2.5% or 2mm. CONCLUSIONS The accuracy of the algorithm was found to be satisfactory at 6 and 10MV both in homogeneous and inhomogeneous situations and in the simulated treatment plans. The algorithm was more accurate than the Pencil Beam Convolution model, particularly in the presence of low density heterogeneities.
منابع مشابه
Point Dose Measurement for Verification of Treatment Planning System using an Indigenous Heterogeneous Pelvis Phantom for Clarkson, Convolution, Superposition, and Fast Superposition Algorithms
Background: Nowadays, advanced radiotherapy equipment includes algorithms to calculate dose. The verification of the calculated doses is important to achieve accurate results. Mostly homogeneous dosimetric phantoms are available commercially which do not mimic the actual patient anatomy; therefore, an indigenous heterogeneous pelvic phantom mimicking actual human pelvic region has been used to ...
متن کاملClinical implications of the anisotropic analytical algorithm for IMRT treatment planning and verification.
PURPOSE To determine the implications of the use of the Anisotropic Analytical Algorithm (AAA) for the production and dosimetric verification of IMRT plans for treatments of the prostate, parotid, nasopharynx and lung. METHODS 72 IMRT treatment plans produced using the Pencil Beam Convolution (PBC) algorithm were recalculated using the AAA and the dose distributions compared. Twenty-four of t...
متن کاملDosimetric Study of an Indigenous and Heterogeneous Pelvic Phantom for Radiotherapy Quality Assurance
Introduction: In vitro dosimetric verification prior to patient treatment plays a key role in accurate and precision radiotherapy treatment delivery. Since the human body is a heterogeneous medium, the aim of this study was to design a heterogeneous pelvic phantom for radiotherapy quality assurance. Material and Methods: A pelvic phantom was ...
متن کاملDosimetric accuracy of the Acuros XB and Anisotropic analytical algorithm near interface of the different density media for the small fields of a 6- MV flattening-filter-free beam
Background: This study was conducted to assess the accuracy of dose calculation near the air-phantom interface of a heterogeneous phantom for Acuros XB (AXB) and Anisotropic Analytical Algorithm (AAA) algorithm of a 6-MV flattening-filter-free beam, compared with film measurements. Materials and Methods: A phantom included air gap was ...
متن کاملdimensional film dosimetry with GAFCHROMIC films for quality assurance and dosimetric verification of 3D conformal radiotherapy in the presence of heterogeneities
Introduction: The presence of heterogeneities such as air-filled cavities in the head and neck treatment fields region, may result in potential dosimetric disagreement because the losses of charged particle equilibrium. Most of treatments planning systems are not able to predict dose distribution of inhomogeneities region accurately. Therefore, dose calculation algorithms need to...