Evaluation of linear measurements of implant sites based on head orientation during acquisition: An ex vivo study using cone-beam computed tomography
نویسندگان
چکیده
PURPOSE This study evaluated the effect of various head orientations during cone-beam computed tomography (CBCT) image acquisition on linear measurements of potential implant sites. MATERIALS AND METHODS Six dry human skulls with a total of 28 implant sites were evaluated for seven different head orientations. The scans were acquired using a Hitachi CB-MercuRay CBCT machine. The scanned volumes were reconstructed. Horizontal and vertical measurements were made and were compared to measurements made after simulating the head position to corrected head angulations. Data was analyzed using a two-way ANOVA test. RESULTS Statistical analysis revealed a significant interaction between the mean errors in vertical measurements with a marked difference observed at the extension head position (P<0.05). Statistical analysis failed to yield any significant interaction between the mean errors in horizontal measurements at various head positions. CONCLUSION Head orientation could significantly affect the vertical measurements in CBCT scans. The main head position influencing the measurements is extension.
منابع مشابه
The Accuracy of Cone-beam Computerized Tomography Linear Measurements in Human Dry Mandible
Background & Aims: The aim of this study was to evaluate the accuracy of Cone-beam Computerized Tomography (CBCT) linear measurements in dry mandibles by comparing them with direct measurements. Methods: Two human dry mandibles were used in this study. Thirteen sites were selected on each mandible. Specimens were scanned by CBCT and linear measurements were made using Planmeca Romexis 2.3.0.R s...
متن کاملEvaluation of prevalence and pattern of maxillary sinus mucosal thickening in cone beam computed tomography of dental implant candidates in Shiraz, Iran
BACKGROUND AND AIM: This study aimed to assess maxillary sinus floor membrane thickness and ostium patency (OP) in patients undergoing cone beam computed tomography (CBCT) evaluation for implant placement in the posterior maxilla. Mucosal thickening (MT) could be cause of post-operative complications in implant procedures; awareness of this incidence could help proper detection and further succ...
متن کاملDiagnostic efficacy of a modified low-dose acquisition protocol for the preoperative evaluation of mini-implant sites
PURPOSE The objective of this study was to compare the outcomes of surgical mini-implant placement when potential mini-implant sites were scanned using a lower-dose 180° acquisition protocol versus a conventional 360° acquisition protocol. MATERIALS AND METHODS Ten dentate human skulls were used to provide sites for potential mini-implant placement. The sites were randomly divided into 2 grou...
متن کاملEvaluation of Metal Artifacts in Cone Beam Computed Tomography by Metal Supported Porcelain Crowns Using Different FOV and Localizations: An In Vitro Study
Introduction: Metal-supported porcelain crowns (MSPC) and bridge restorations may be present in the mouths of patients undergoing CBCT imaging. Artifacts that are caused by these MSPCs may adversely affect image quality. The aim of this study is to determine the effect of different FOV (field of view) and localization in FOV on metal artifacts caused by MSPC. Methods:</...
متن کاملEffect of Slice Thickness on the Accuracy of Linear Measurements Made on Cone Beam Computed Tomography Images (InVitro
Objective: Cone beam computed tomography (CBCT) is applied for the imaging of the maxillofacial and dental structures, particularly for surgical treatments and dental implants. The aim of this study was to assess the effect of slice thickness on the accuracy of CBCT linear measurements. Methods: In this diagnostic accuracy study, forty-two titanium pins with the same dimensions were inserted...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 45 شماره
صفحات -
تاریخ انتشار 2015