Improved fluorescence polarization assay for use in evaluating fetal lung maturity. I. Development of the assay procedure.
نویسندگان
چکیده
We describe a fluorescence polarization assay for use in predicting fetal lung maturity, which is suitable for the TDx Analyzer (an automated fluorescence polarimeter). The assay requires 0.5 mL of amniotic fluid and approximately 30 min, and involves the fluorophore 1-palmitoyl-2(6-[(7-nitro-2, 1, 3-benzoxadiazol-4-yl)amino]caproyl) phosphatidylcholine. Reproducible polarization measurements depend on proper regulation of incubation time and temperature, but variations in the concentrations of fluorescent probe and amniotic fluid have little effect on measured polarization and therefore little effect on assay precision. Working solutions of the fluorescent probe are stable for at least nine months when stored at -20 degrees C and pH 5. Interferences include erythrocytes, serum, bilirubin, meconium, and lidocaine.
منابع مشابه
Improved fluorescence polarization assay for use in evaluating fetal lung maturity. II. Analytical evaluation and comparison with the lecithin/sphingomyelin ratio.
We assessed the analytical performance of an improved fluorescence polarization assay for use in evaluating fetal lung maturity and compared results with the lecithin/sphingomyelin ratio. During a three-month period 150 patients' samples were assayed by clinical laboratory personnel with both techniques. Values for the lecithin/sphingomyelin ratio correlate closely with net fluorescence polariz...
متن کاملProspective clinical evaluation of an improved fluorescence polarization assay for predicting fetal lung maturity.
We performed a prospective clinical evaluation of our newly developed fluorescence polarization procedure to predict fetal lung maturity (Clin Chem 1986;32:248-54). Net fluorescence polarization was measured at 34 degrees C after a 6.5-min incubation of amniotic fluid with fluorophore. For the 26 cases of neonatal respiratory distress syndrome encountered in 196 deliveries, the net polarization...
متن کاملDegree of unsaturation of the fatty acid chains of phospholipids influences the fluorescence polarization: implications for evaluating fetal lung maturity.
To study the effect of fatty acid chain saturation on the fluorescence polarization assay as a measure of fetal lung maturity, we used purified phospholipids isolated from human amniotic fluid and various commercial phospholipids. We found that the fluorescence polarization value decreased as the concentration of unsaturated fatty acids increased. In contrast, the lecithin/sphingomyelin ratio i...
متن کاملA calibrated fluorescence polarization assay for assessment of fetal lung maturity.
The mechanism of a new fluorescence polarization method for measuring phospholipid in amniotic fluid is described. The fluorescence polarization of a surfactant fluorescent dye solubilized in amniotic fluid correlates significantly with the L/S ratio. Model studies indicate that this results primarily from partitioning of the dye between endogenous albumin, which causes a high fluorescence pola...
متن کاملAmniotic fluid polarization of fluorescence and lecithin/sphingomyelin ratio decision criteria assessed.
A negative finding of amniotic fluid (AF) phosphatidyl glycerol (PG) does not eliminate the need for determining the lecithin/sphingomyelin ratio (LSR). We use a novel approach to classify fetal lung maturity (FLM) data, and to validate the fluorescence polarization (FP) surfactant assay (Abbott), which replaces the PG assay and reduces the frequency of repeat LSR. This method finds the values ...