Dynamic contour tonometry: a comparative study on human cadaver eyes.

نویسندگان

  • Christoph Kniestedt
  • Michelle Nee
  • Robert L Stamper
چکیده

OBJECTIVE To compare intraocular pressure measurements obtained by recently introduced dynamic contour tonometry (DCT), Goldmann applanation tonometry (GAT), pneumatonometry (PTG), and intracameral manometry in human cadaver eyes. METHODS Sixteen freshly enucleated human cadaver eyes were deepithelialized and dehydrated with dextran. A tube was placed in the anterior chamber and connected to a transducer and to a bottle system filled with balanced salt solution. The pressure in the eye was then altered between 5 mm Hg and 58 mm Hg. Intraocular pressure measurements were obtained with DCT, GAT, and PTG at each manometric pressure reading. RESULTS On average, DCT values measured 0.58 +/- 0.70 mm Hg higher than real intracameral pressure. The GAT and PGT showed consistently lower values, -4.01 +/- 1.76 mm Hg and -5.09 +/- 2.61 mm Hg, respectively. At all bottle heights, DCT values were significantly closer to the reference pressure than GAT and PTG (P<.001). CONCLUSIONS Measurement with DCT provides IOP values significantly closer to true manometric levels than either GAT or PTG. Further studies are warranted to determine its reliability in patients and the effect of corneal thickness.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dynamic contour tonometry and goldman applanation tonometry in eyes with keratoconus.

PURPOSE The dynamic contour tonometer is a nonapplanation contact tonometer designed to be largely independent of the structural properties of the cornea. Theoretically, it may measure intraocular pressure most accurately in abnormally thinner corneas. This study compares intraocular pressure measurements by dynamic contour tonometry with Goldman applanation tonometry in eyes with normal cornea...

متن کامل

Correlation between diurnal variation of intraocular pressure, ocular pulse amplitude and corneal structural properties.

PURPOSE The purpose of this study was to evaluate the fluctuation of intraocular pressure measurements obtained by Goldmann applanation tonometry, dynamic contour tonometry, and corneal compensated non-contact tonometry during office hours in glaucoma and healthy participants. This study also aims at correlating the intraocular pressure fluctuations with fluctuations of corneal hysteresis, cent...

متن کامل

Intraocular pressure measurements with Goldmann applanation tonometry and dynamic contour tonometry in eyes after IntraLASIK or LASEK

BACKGROUND AND METHODS Myopic photorefractive surgery induces a reduction in central corneal thickness, which may lead to underestimation of intraocular pressure. This retrospective clinical study compared intraocular pressure measurements obtained by Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT-Pascal) in eyes undergoing myopic intralaser-assisted in situ keratomileu...

متن کامل

Comparison Between Dynamic Contour Tonometry and Goldmann Applanation Tonometry

PURPOSE To compare the intraocular pressures (IOPs) measured by dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT), and to investigate the association of IOPs on eyes of varying central corneal thickness (CCT). METHODS In this prospective study, 451 eyes of 233 subjects were enrolled. IOPs were measured by GAT and DCT. CCT was measured three times and the average was cal...

متن کامل

Applanation tonometry versus dynamic contour tonometry in eyes treated with latanoprost.

PURPOSE To examine the differences between Goldmann Applanation Tonometry (GAT) and Dynamic Contour Tonometry (DCT) associated with latanoprost use. METHODS Twenty-four eyes (of 24 patients) treated with latanoprost monotherapy (latanoprost group, LG), 11 eyes (of 11 patients) not receiving prostaglandin analogs (nonlatanoprost group, NLG), and 20 eyes of 20 nonglaucomatous patients (control ...