Volume-of-interest cone-beam CT using a 2.35 MV beam generated with a carbon target.
نویسندگان
چکیده
PURPOSE This is a proof-of-concept study addressing volume of interest (VOI) cone beam CT (CBCT) imaging using an x-ray beam produced by 2.35 MeV electrons incident on a carbon linear accelerator target. Methodology is presented relevant to VOI CBCT image acquisition and reconstruction. Sample image data are given to demonstrate and compare two approaches to minimizing artifacts arising from reconstruction with truncated projections. Dosimetric measurements quantify the potential dose reduction of VOI acquisition relative to full-field CBCT. The dependence of contrast-to-noise ratio (CNR) on VOI dimension is investigated. METHODS A paradigm is presented linking the treatment planning process with the imaging technique, allowing definition of an imaging VOI to be tailored to the geometry of the patient. Missing data in truncated projection images are completed using a priori information in the form of digitally reconstructed radiographs (DRRs) generated from the planning CT set. This method is compared to a simpler technique of extrapolating truncated projection data prior to reconstruction. The utility of these approaches is shown through imaging of a geometric phantom and the head-and-neck section of a lamb. The total scatter factor of the 2.35 MV∕carbon beam on field size is measured and compared to a standard therapeutic beam to estimate the comparative dose reduction inside the VOI. Thermoluminescent dosimeters and Gafchromic film measurements are used to compare the imaging dose distributions for the 2.35 MV∕carbon beam between VOI and full-field techniques. The dependence of CNR on VOI dimension is measured for VOIs ranging from 4 to 15 cm diameter. RESULTS Without compensating for missing data outside of truncated projections prior to reconstruction, pronounced boundary artifacts are present, in three dimensions, within 2-3 cm of the edges of the VOI. These artifacts, as well as cupping inside the VOI, can be reduced substantially using either the DRR filling or extrapolation techniques presented. Compared to 6 MV, the 2.35 MV∕carbon beam shows a substantially greater dependence of total scatter factor on field size, indicating a comparative advantage of the VOI approach when combined with the low-Z target beam. Dosimetric measurements in the anthropomorphic head phantom demonstrate a dose reduction by up to 15% and 75% inside and outside of the VOI, respectively, compared to full-field imaging. For the 2.35 MV∕carbon beam, CNR was shown to be approximately invariant with VOI dimension for bone and lung objects. CONCLUSIONS The low-Z target, VOI CBCT technique appears to be feasible and combines the desirable characteristics of the low-Z target beam with regard to CNR, with the capacity to localize the imaging dose to the anatomy relevant to the image guidance task.
منابع مشابه
A feasibility study on the use of MV-CBCT images for urgent palliative treatment planning
Introduction: The application of 3D volumetric imaging modalities in treatment planning of radiation therapy can provide more precisely define tumor localization, and computed tomography (CT) is the most common accepted method for treatment planning. Given the lack of a CT scanner stationed in all radiotherapy departments and equipping most of the medical linear accelerators wi...
متن کاملCalculation and Comparison of Heart Integral Dose in The Treatment of Esophagus Cancer with Three Photon Energies & Using CT Simulation and Treatment Planning System
Introduction: Esophageal cancer is one of the most frequently occurring cancers in Iran and having a high incidence rate among other countries. Radiotherapy is one of the three methods (surgery, radiotherapy and chemotherapy) for radical or palliative treatment of esophageal cancer. In this method of treatment, the organs such as heart and spinal cord are regarded as organs at r...
متن کاملCorrelation Between Tissue Densities in Computed Tomography and Three Different Cone-Beam Computed Tomography Units (In Vitro)
Background and Aim: Bone density is of great assistance in the selection of the proper implant site. The present study aimed to assess the correlation between tissue densities in computed tomography (CT) and three different cone-beam computed tomography (CBCT) units. Materials and Methods: In this descriptive study, a radiographic phantom consisting of a transparent polymethyl methacrylate (PM...
متن کاملThe Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography
Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...
متن کاملMega Voltage Cone Beam Computed Tomography (MV- CBCT) using a Standard Medical Linear Accelerator and EPID: A feasibility study.
Introduction: The success of radiotherapy cancer treatment delivery depends on the accuracy of patient positioning for each treatment session. A number of kilovoltage x-ray volumetric imaging modalities with an additional source and detector have been developed to allow patient set-up verification based on the internal anatomy, but a significant portion of medical linacs are on...