Scattered radiation doses absorbed by technicians at different distances from X-ray exposure: Experiments on prosthesis.
نویسندگان
چکیده
UNLABELLED This work aimed to investigate the spatial distribution of scattered radiation doses induced by exposure to the portable X-ray, the C-arm machine, and to simulate the radiologist without a shield of lead clothing, radiation doses absorbed by medical staff at 2 m from the central exposure point. MATERIAL AND METHOD With the adoption of the Rando Phantom, several frequently X-rayed body parts were exposed to X-ray radiation, and the scattered radiation doses were measured by ionization chamber dosimeters at various angles from the patient. Assuming that the central point of the X-ray was located at the belly button, five detection points were distributed in the operation room at 1 m above the ground and 1-2 m from the central point horizontally. RESULTS The radiation dose measured at point B was the lowest, and the scattered radiation dose absorbed by the prosthesis from the X-ray's vertical projection was 0.07 ±0.03 μGy, which was less than the background radiation levels. The Fluke biomedical model 660-5DE (400 cc) and 660-3DE (4 cc) ion chambers were used to detect air dose at a distance of approximately two meters from the central point. The AP projection radiation doses at point B was the lowest (0.07±0.03 μGy) and the radiation doses at point D was the highest (0.26±0.08 μGy) .Only taking the vertical projection into account, the radiation doses at point B was the lowest (0.52 μGy), and the radiation doses at point E was the highest (4 μGy).The PA projection radiation at point B was the lowest (0.36 μGy) and the radiation doses at point E was the highest(2.77 μGy), occupying 10-32% of the maximum doses. The maximum dose in five directions was nine times to the minimum dose. When the PX and the C-arm machine were used, the radiation doses at a distance of 2 m were attenuated to the background radiation level. The radiologist without a lead shield should stand at point B of patient's feet. Accordingly, teaching materials on radiation safety for radiological interns and clinical technicians were formulated.
منابع مشابه
Dosimetry of Occupational Radiation around Panoramic X-ray Apparatus
Background: Panoramic imaging is one of the most common imaging methods in dentistry. Regarding the side-effects of ionizing radiation, it is necessary to survey different aspects and details of panoramic imaging. In this study, we compared the absorbed x-ray dose around two panoramic x-ray units: PM 2002 CC Proline (Planmeca, Helsinki, Finland) and Cranex Tome (Soredex, Helsinki, Finland).<br ...
متن کاملRetrospective assessment of radiation dose by Fluorescence In Situ Hybridization and evaluation of stable chromosomal aberrations
Estimation of absorbed dose for radiation workers or person involved in different radiological accident is the aim of biodosimetry. Cytogenetic methods are the most current and applicable biodosimetry tools. In chronic or protracted exposure, fluorescence in situ hybridization (FISH), stable chromosomal aberration is used for estimation of absorbed dose. For precise estimation of absorbed dose,...
متن کاملImpact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals
ABSTRACT Background: The X-ray machines used for radiodiagnosis should meet certain quality assurance (QA) programmes . These are necessary to have good quality radiographs at reasonably low exposure to patients. Materials and Methods: Dose reduction methods in abdomen X-ray examination were carried out in 10 hospitals in Tehran. This paper presents the work, which was implemented on 200 ...
متن کاملMonte Carlo radiography simulation for assessment of absorbed radiation dose in femur bone marrow during X-ray radiography for constant mAs and AEC techniques
Background: The purpose of this study was to find the accurate absorbed dose in the femur bone marrow during the X-ray radiography for constant mAs and AEC techniques. Materials and Methods: The DOSXYZnrc was used to simulate radiation doses in two human femurs during diagnostic radiography. EGSnrc phantoms produced from actual CT images of human femurs were modified by adding seven micrometre ...
متن کاملبررسی میزان پرتو پراکنده Xو دوز دریافتی کارکنان بخش آنژیوگرافی یک بیمارستان نظامی
Background and aims: Using X-ray for diagnostic test, is one of the most important exposure to radiation sources for general population and in particular for radiation workers. Therefore, holding down the pollution at the lowest possible level is very important for protection against the internal exposure. The objective of this study is to survey the scattered X-ray rate and received dose by st...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Bio-medical materials and engineering
دوره 26 Suppl 1 شماره
صفحات -
تاریخ انتشار 2015