Extended clinical trial and evaluation of urea nitrogen determination with the Ektachem GLU/BUN analyzer.

نویسندگان

  • Z L Bandi
  • J B Fuller
  • D E Bee
  • G P James
چکیده

We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analysis for urea nitrogen with the multilayer film analysis system ("Ektachem"). We analyzed 456 clinical samples with this instrument, by the manual urease/glucose dehydrogenase method, and with the Beckman System I GLU/BUN Analyzer. Precision and accuracy were estimated for 50, 220, 270, and 500 mg/L urea nitrogen concentrations in 100, 30, or 20 microL of serum. Potential interference of 15 compounds was evaluated. Random error (defined as 1.965 X SD) was 7, 10, 12, and 18 mg/L. Systematic error was 3, 4, 5, and 15 mg/L. Total analytical error was 11, 14, 17, and 34 mg/L for analysis of 100 microL of serum at the above-stated urea nitrogen concentrations. The greatest interference (6 mg/L) was caused by ethanol (300 mg/L) and by hemoglobin (500 mg/L) in the urea nitrogen (at 260 mg/L) determination. Urea nitrogen concentration, as determined with the Ektachem was linearly related to the expected concentration, at least up to 1187 mg/L. Carryover was not statistically significant.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Extended clinical trial and evaluation of glucose determination with the Eastman Kodak Ektachem GLU/BUN Analyzer.

We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analyses for glucose with the "Ektachem." We analyzed 760 clinical samples by this technique, by the FDA Proposed Class Standard glucose reference method, and with the Beckman System I GLU/BUN Analyzer. Precision and accuracy we...

متن کامل

Evaluation of an engineering model of the "EKTACHEM" analyzer for glucose and urea assay.

We report results of a 21-week evaluation of an engineering model of the Kodak “EKTACHEM Glu/BUN Analyzer.” This novel clinical instrument measures glucose and “urea nitrogen” (actually, urea) in a minimum of 30 j.tL of body fluids by use of the principles of thin, multilayer film analysis and reflectance spectrophotometry. We compared the performance of the instrument to that of the Dupont aca...

متن کامل

Abstracts/Sommaires/Zusammenfassungen

An evaluation of the Kodak glucose/BUN analyzer including experience with proposed testing protocols An evaluation of the analysis of serum urea and plasma_glucose on the Kodak Ektachem GLU/BUN()analyzer is described. It incorporates the experimental design and statistical methods outlined in the Proposed Standard for Establishing Performance Claims (PSEP) (1) for clinical chemical methods PSEP...

متن کامل

Dipyrone interference on several common biochemical tests.

We studied the effect in vitro and in vivo of dipyrone on the determination of several biochemical tests in two analyzers, a Hitachi 747 and a Kodak Ektachem 700. From studies in vitro, we found significant interference by dipyrone (P < 0.05) in the determination of creatine kinase (CK), lactate dehydrogenase (LD), uric acid, triglycerides, cholesterol, aspartate aminotransferase, alanine amino...

متن کامل

Applicability of the Ektachem 400 analyzer for assaying analytes in miscellaneous body fluids.

The Ektachem analyzer, a slide-film chemistry technology, demonstrated applicability to non-serum, non-urine stimulated and actual clinical specimens having a wide range of protein concentration and pH. The simulated specimens studied had pH and protein that ranged from 2 to 9 and 0 to 88 g/L, respectively. Actual patients' specimens studied, which had protein concentrations ranging from 0.1 to...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 27 3  شماره 

صفحات  -

تاریخ انتشار 1981