ATP measurement as method to monitor the quality of reprocessing flexible endoscopes

نویسندگان

  • Dorothea Hansen
  • Daniel Benner
  • Martin Hilgenhöner
  • Therese Leisebein
  • Andreas Brauksiepe
  • Walter Popp
چکیده

Insufficient performance of cleaning and disinfection of flexible endoscopes can pose an infection risk to patients. Actually quality of reprocessing is checked by performing microbiological cultures. Unfortunately, their results are not available on the same day so that more rapid methods are desirable. We compared the ATP (adenosine triphosphate) bioluminescence for hygiene checking of the reprocessing procedures of 108 flexible endoscopes with routine microbiological culture technics. Sensitivity and specifity of ATP bioluminescence was calculated. 28 endoscopes showed bacterial growth of at least one sample. Depending on the applied threshold of bioluminescence between 67 and 28 endoscopes were positive. Sensitivity varied between 0.46 and 0.75 and specifity between 0.43 and 0.81. ATP bioluminescence does not replace routine microbiologic methods but it can indicate the need of immediate check of reprocessing.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monitoring of endoscope reprocessing with an adenosine triphosphate (ATP) bioluminescence method

Background: The arising challenges over endoscope reprocessing quality proposes to look for possibilities to measure and control the process of endoscope reprocessing. Aim: The goal of this study was to evaluate the feasibility of monitoring endoscope reprocessing with an adenosine triphosphate (ATP) based bioluminescence system. Methods: 60 samples of eight gastroscopes have been assessed from...

متن کامل

Computer-aided recording of automatic endoscope washing and disinfection processes as an integral part of medical documentation for quality assurance purposes

BACKGROUND The reprocessing of medical endoscopes is carried out using automatic cleaning and disinfection machines. The documentation and archiving of records of properly conducted reprocessing procedures is the last and increasingly important part of the reprocessing cycle for flexible endoscopes. METHODS This report describes a new computer program designed to monitor and document the auto...

متن کامل

Recent Update on Microbiological Monitoring of Gastrointestinal Endoscopes after High-Level Disinfection

Endoscopy-related infections are important contributors to nosocomial infections. Endoscope reprocessing according to standard guidelines ensures high-level disinfection and prevents endoscopy-related infections. Microbiological surveillance may help in monitoring the effectiveness of gastrointestinal endoscope disinfection. The process involves microbial cultures and non-culture methods such a...

متن کامل

Superbugs on Duodenoscopes: the Challenge of Cleaning and Disinfection of Reusable Devices.

Inadequate flexible endoscope reprocessing has been associated with infection outbreaks, most recently caused by carbapenem-resistant Enterobacteriaceae. Lapses in essential device reprocessing steps such as cleaning, disinfection/sterilization, and storage have been reported, but some outbreaks have occurred despite claimed adherence to established guidelines. Recommended changes in these guid...

متن کامل

Residual bioburden in reprocessed side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP)

BACKGROUND AND STUDY AIM Worldwide some endoscopy units routinely continue to use manual reprocessing techniques for disinfection of side-view endoscopes. The aim of this study was to evaluate the outcome quality of manual reprocessing techniques for removal and inactivation of the bioburden from side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP) in a tertiary r...