Early- and long-term intravascular ultrasound and angiographic findings after bioabsorbable magnesium stent implantation in human coronary arteries.

نویسندگان

  • Ron Waksman
  • Raimund Erbel
  • Carlo Di Mario
  • Jozef Bartunek
  • Bernard de Bruyne
  • Franz R Eberli
  • Paul Erne
  • Michael Haude
  • Mark Horrigan
  • Charles Ilsley
  • Dirk Böse
  • Hans Bonnier
  • Jacques Koolen
  • Thomas F Lüscher
  • Neil J Weissman
چکیده

OBJECTIVES This study aimed to evaluate the degradation rate and long-term vascular responses to the absorbable metal stent (AMS). BACKGROUND The AMS demonstrated feasibility and safety at 4 months in human coronary arteries. METHODS The PROGRESS-AMS (Clinical Performance and Angiographic Results of Coronary Stenting) was a prospective, multicenter clinical trial of 63 patients with coronary artery disease who underwent AMS implantation. Angiography and intravascular ultrasound (IVUS) were conducted immediately after AMS deployment and at 4 months. Eight patients who did not require repeat revascularization at 4 months underwent late angiographic and IVUS follow-up from 12 to 28 months. RESULTS The AMS was well-expanded upon deployment without immediate recoil. The major contributors for restenosis as detected by IVUS at 4 months were: decrease of external elastic membrane volume (42%), extra-stent neointima (13%), and intra-stent neointima (45%). From 4 months to late follow-up, paired IVUS analysis demonstrated complete stent degradation with durability of the 4-month IVUS indexes. The neointima was reduced by 3.6 +/- 5.2 mm(3), with an increase in the stent cross sectional area of 0.5 +/- 1.0 mm(2) (p = NS). The median in-stent minimal lumen diameter was increased from 1.87 to 2.17 mm at long-term follow-up. The median angiographic late loss was reduced from 0.62 to 0.40 mm by quantitative coronary angiography from 4 months to late follow-up. CONCLUSIONS Intravascular ultrasound imaging supports the safety profile of AMS with degradation at 4 months and maintains durability of the results without any early or late adverse findings. Slower degradation is warranted to provide sufficient radial force to improve long-term patency rates of the AMS.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Acute Recoil between Bioabsorbable Poly-L-lactic Acid XINSORB Stent and Metallic Stent in Porcine Model

OBJECTIVE To investigate acute recoil of bioabsorbable poly-L-lactic acid (PLLA) stent. BACKGROUND As newly developed coronary stent, bioabsorbable PLLA stent still encountered concern of acute stent recoil. METHODS Sixteen minipigs were enrolled in our study. Eight PLLA XINSORB stents (Weite Biotechnology Co., Ltd., China) and eight metallic stents (EXCEL, Jiwei Co., Ltd. China) were impla...

متن کامل

Relationship between neointimal thickness and shear stress after Wallstent implantation in human coronary arteries.

BACKGROUND In-stent restenosis by excessive intimal hyperplasia reduces the long-term clinical efficacy of coronary stents. Because shear stress (SS) is related to plaque growth in atherosclerosis, we investigated whether variations in SS distribution are related to variations in neointima formation. METHODS AND RESULTS In 14 patients, at 6-month follow-up after coronary Wallstent implantatio...

متن کامل

Persistent inhibition of neointimal hyperplasia after sirolimus-eluting stent implantation: long-term (up to 2 years) clinical, angiographic, and intravascular ultrasound follow-up.

BACKGROUND Early results of sirolimus-eluting stent implantation showed a nearly complete abolition of neointimal hyperplasia. The question remains, however, whether the early promising results will still be evident at long-term follow-up. The objective of our study was to evaluate the efficiency of sirolimus-eluting stent implantation for up to 2 years of follow-up. METHODS AND RESULTS Fifte...

متن کامل

Stent design related neointimal tissue proliferation in human coronary arteries; an intravascular ultrasound study.

AIMS Histological restenosis models in animals have indicated that stent design has a significant impact on vessel trauma during stent implantation and on the amount of subsequent neointimal tissue proliferation. The impact of different stent designs on intimal hyperplasia in human atherosclerotic coronary arteries has not been determined. METHODS AND RESULTS Angiographic and intravascular ul...

متن کامل

Acute and Long Term Outcomes of Coronary Intervention in Unprotected Left Main Lesions

Introduction : Significant left main coronary artery stenosis jeopardizes the entire myocardium of the left ventricle and has the worst prognosis of any form of coronary artery disease. Coronary-artery bypass grafting (CABG) has been considered as the standard therapeutic approach for such patients. There are limited data on the safety and effectiveness of percutaneous coronary intervention (PC...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • JACC. Cardiovascular interventions

دوره 2 4  شماره 

صفحات  -

تاریخ انتشار 2009