The thermal effects of lavage on 57 ox femoral heads prepared for hip resurfacing arthroplasty

نویسندگان

  • Richard P Baker
  • Michael R Whitehouse
  • Angus Maclean
  • Ashley W Blom
  • Gordon C Bannister
چکیده

BACKGROUND AND PURPOSE Previously, we have documented surface temperatures recorded by thermography great enough to cause osteonecrosis of the femoral head during hip resurfacing. We now performed an in vitro investigation with 3 questions: (1) whether water irrigation reduced bone surface temperature, (2) whether external bone temperatures were similar to core temperatures, and (3) whether blunting of the reamer affected temperature generation. METHODS Using an ox-bone model, 57 femoral heads were peripherally reamed. The surface temperatures of bone were measured using a thermal camera and internal bone temperatures were measured using 2 theromocouples. We measured the effects of cooling with water at room temperature and with ice-cooled water. Progressive blunting of reamers was assessed over the 57 experiments. RESULTS Mean and maximum temperatures generated during peripheral reaming were greater when no irrigation was used. Ice-cold saline protected femoral heads from thermal damage. External bone temperatures were much greater than internal temperatures, which were not sufficiently elevated to cause osteonecrosis regardless of lavage. Blunting of the reamer was not found to have a statistically significant effect in this study. INTERPRETATION Cooling with ice-cooled water is recommended. Internal bone temperatures are not elevated despite the high surface temperatures reached during femoral head resurfacing.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Maximum temperatures of 89°C recorded during the mechanical preparation of 35 femoral heads for resurfacing

BACKGROUND AND PURPOSE We noticed that our instruments were often too hot to touch after preparing the femoral head for resurfacing, and questioned whether the heat generated could exceed temperatures known to cause osteonecrosis. PATIENTS AND METHODS Using an infra-red thermal imaging camera, we measured real-time femoral head temperatures during femoral head reaming in 35 patients undergoin...

متن کامل

Reduction of the potential for thermal damage during hip resurfacing.

Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. ...

متن کامل

The Vascular Evaluation of Femoral Head Remnant after Resurfacing Hip Arthroplasty

RESURFACING HIP ARTHROPLASTY +*Hananouchi, T; *Sugano, N; *Nishii, T; *Sakai, T; **Miki, H;**Lee, SB; **Ohzono, K; *Yoshikawa, H +*Department of Orthopaedic Surgery, Osaka University Graduate School of Medicine [email protected] Introduction: Metal on metal resurfacing hip arthroplasty (RHA) has been getting popularity due to its concept of bone conservation, stability, and expected longevi...

متن کامل

Notching of the femoral neck during resurfacing arthroplasty of the hip: a vascular study.

During hip resurfacing arthroplasty, excessive valgus positioning or surgical technique can result in notching of the femoral neck. Although mechanical weakening and subsequent fracture of the femoral neck are well described, the potential damage to the retinacular vessels leading to an ischaemic event is relatively unknown. Using laser Doppler flowmetry, we measured the blood flow in 14 osteoa...

متن کامل

Range of motion after stemmed total hip arthroplasty and hip resurfacing - a clinical study.

Range of motion after total hip arthroplasty is becoming an important topic as today's patients present at a younger age and are more active. An effective study design to carry out comparisons of clinical performance between two implants should eliminate patient-related extraneous variables (e.g., age, gender, activity level, among others). The aim of the present study was to compare the range ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 84  شماره 

صفحات  -

تاریخ انتشار 2013