The aim of this study was to evaluate complications experienced during implantation of the Braile Vena Cava filter (VCF) and the efficacy of the centralization mechanism of the filter. This retrospective cohort study evaluated all Braile Biomédica VCFs implanted from 2004 to 2009 in Hospital de Base Medicine School in São José do Rio Preto, Brazil. Of particular concern was the filter's symmetry during implantation and complications experienced during the procedure. All the angiographic examinations performed during the implantation of the filters were analyzed in respect to the following parameters: migration of the filter, non-opening or difficulties in the implantation and centralization of the filter. A total of 112 Braile CVFs were implanted and there were no reports of filter opening difficulties or in respect to migration. Asymmetry was observed in 1/112 (0.9%) cases. A statistically significant difference was seen on comparing historical data on decentralization of the Greenfield filter with the data of this study. The Braile Biomédico filter is an evolution of the Greenfield filter providing improved embolus capture and better implantation symmetry.
منابع مشابه
Strut Deformation of the Greenfield IVC Filter Associated with Blunt Trauma
We report an unusual traumatic complication of inferior vena cava (IVC) perforation by a Greenfield filter (GF) in a 19-year-old woman. She was inserted the Greenfield IVC filter in suprarenal position because of deep vein thrombosis associated with pulmonary embolism. Six months later, after repeated falls during skiing, she was readmitted to our hospital for paralytic ileus. A plain abdominal...
متن کاملImpact of unilateral common iliac vein occlusion on trapping efficacy of the Greenfield filter: an in vitro study.
RATIONALE AND OBJECTIVES The purpose of this study was to assess the effect of unilateral common iliac vein occlusion on the capturing efficacy of the Greenfield filter in vitro. MATERIALS AND METHODS A stainless steel over-the-wire Greenfield filter was placed in the Silastic inferior vena cava module of a pulsatile circuit. Three 30-mm blood clots in sets of five were injected through the m...
متن کاملVena caval filter use in patients with sepsis: results in 175 patients.
BACKGROUND Septic patients are at risk of thromboembolism. However, the Food and Drug Administration guidance for intravascular filters states that "filters should not be implanted in patients with risk of septic embolism." The purpose of this study is to evaluate this restriction. HYPOTHESIS There is no difference in outcomes following filter placement in patients with and without septicemia...
متن کاملRenal vein thrombosis with recurrent pulmonary emboli in the nephrotic syndrome: use of the Greenfield filter.
We describe a case of renal vein thrombosis in association with the nephrotic syndrome. The patient had recurrent pulmonary emboli which, having failed to respond to routine anticoagulant treatment, were managed successfully by the insertion of a Greenfield filter into the inferior vena cava above the renal vein thrombosis site. The only coagulation abnormality in this patient was a deficiency ...
متن کاملInferior Vena Cava Filters, May-Thurner Syndrome, and Vein Stents.
IVC Filters What Is an IVC Filter? What Is Its Purpose? Patients who have a blood clot in their leg, referred to as deep vein thrombosis (DVT), are at risk of the clot breaking off and traveling toward the lung. It gets to the lung via the big vein in the abdomen (Figure 1), called the inferior vena cava (IVC). The traveling clot is called an embolus. Once it reaches the lung (after having pass...