Subcutaneous Tissue Thickness is an Independent Predictor of Image Noise in Cardiac CT

نویسندگان

  • Henrique Lane Staniak
  • Rodolfo Sharovsky
  • Alexandre Costa Pereira
  • Cláudio Campi de Castro
  • Isabela M. Benseñor
  • Paulo A. Lotufo
  • Márcio Sommer Bittencourt
چکیده

BACKGROUND Few data on the definition of simple robust parameters to predict image noise in cardiac computed tomography (CT) exist. OBJECTIVES To evaluate the value of a simple measure of subcutaneous tissue as a predictor of image noise in cardiac CT. METHODS 86 patients underwent prospective ECG-gated coronary computed tomographic angiography (CTA) and coronary calcium scoring (CAC) with 120 kV and 150 mA. The image quality was objectively measured by the image noise in the aorta in the cardiac CTA, and low noise was defined as noise < 30 HU. The chest anteroposterior diameter and lateral width, the image noise in the aorta and the skin-sternum (SS) thickness were measured as predictors of cardiac CTA noise. The association of the predictors and image noise was performed by using Pearson correlation. RESULTS The mean radiation dose was 3.5 ± 1.5 mSv. The mean image noise in CT was 36.3 ± 8.5 HU, and the mean image noise in non-contrast scan was 17.7 ± 4.4 HU. All predictors were independently associated with cardiac CTA noise. The best predictors were SS thickness, with a correlation of 0.70 (p < 0.001), and noise in the non-contrast images, with a correlation of 0.73 (p < 0.001). When evaluating the ability to predict low image noise, the areas under the ROC curve for the non-contrast noise and for the SS thickness were 0.837 and 0.864, respectively. CONCLUSION Both SS thickness and CAC noise are simple accurate predictors of cardiac CTA image noise. Those parameters can be incorporated in standard CT protocols to adequately adjust radiation exposure.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessing the image quality and eye lens dose reduction using bismuth shielding in computed tomography of brain

Background: Epidemiological studies show that computed tomography (CT) is one of the main sources of ionizing radiations. Shielding of radiosensitive organs is one of the dose reduction methods. This study aimed to assess the eye lens dose reduction and image quality resulting from the use of radio-protective bismuth shield in brain CT imaging. Methods: Bismut...

متن کامل

Evaluation of attenuation correction process in cardiac SPECT images

  Introduction: Attenuation correction is a useful process for improving myocardial perfusion SPECT and is dependent on activity and distribution of attenuation coefficients in the body (attenuation map). Attenuation artifacts are a common problem in myocardial perfusion SPECT. The aim of this study was to compare the effect of attenuation correction using different a...

متن کامل

Investigation of dosimetric characteristic of NIPAM polymer gel using x-ray CT

Introduction: Polymer gel dosimeters contain chemical materials sensitive to the radiation which are polymerized by the radiation as a function of absorbed dose. So information of spatial dose distribution can be extracted by imaging from irradiated gel. Among imaging techniques, computed tomography (CT) poses as an attractive method because of practical advantages such as acce...

متن کامل

Shearlet-Based Adaptive Noise Reduction in CT Images

The noise in reconstructed slices of X-ray Computed Tomography (CT) is of unknown distribution, non-stationary, oriented and difficult to distinguish from main structural information. This requires the development of special post-processing methods based on the local statistical evaluation of the noise component. This paper presents an adaptive method of reducing noise in CT images employing th...

متن کامل

Comparing X-ray dose reduction capability of silicon-bismuth micro- and nanocomposite shields using chest CT test

Comparison of radiation shields of different thicknesses and structures (especially different particle sizes of their constituents) is helpful to select an appropriate shield for patient protection in medical imaging tests particularly CT-scan. In this study micro- and nano-powders of bismuth were mixed with silicone matrix to make composite radiation shields for breast. The shields were provid...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 102  شماره 

صفحات  -

تاریخ انتشار 2014