Magnetically programmable shunt valve: MRI at 3-Tesla.
نویسندگان
چکیده
A magnetically programmable cerebrospinal fluid (CSF) shunt valve (Codman Hakim Programmable Valve, Codman, a Johnson & Johnson Company, Raynham, MA) was assessed for magnetic field interactions, heating, artifacts and functional changes at 3-Tesla. The programmable valve showed minor magnetic field interactions and heating (+0.4 degrees C). Artifacts were relatively large in relation to the size and shape of this implant and, as such, may create a problem if the area of interest is in proximity to this implant. While multiple exposures and various magnetic resonance imaging (MRI) conditions at 3-Tesla changed the settings of some valves (i.e., reprogramming was needed), the function of the programmable valve was not permanently affected. Therefore, this magnetically programmable CSF shunt valve is acceptable for a patient undergoing MRI at 3-Tesla or less when specific safety guidelines are followed, including resetting the valve, as needed.
منابع مشابه
Shunt Devices for the Treatment of Adult Hydrocephalus: Recent Progress and Characteristics
Various types of shunt valves have been developed during the past 50 years, most of which can be classified into the following categories: (1) fixed differential pressure valves; (2) fixed differential pressure (DP) valves with an antisiphon mechanism; (3) programmable DP valves; (4) programmable DP valves with an antisiphon mechanism; and (5) programmable antisiphon valves. When considering th...
متن کاملProgrammable valve shunts: are they really better?
Programmable valve shunts allows selection of opening pressure of shunt valve. In the presented article, a unique complication pertaining to programmable shunts has been discussed. A 5-year-old boy who had tectal plate low grade glioma with obstructive hydrocephalus was managed with Codman programmable ventriculoperitoneal shunt. There was a spontaneous change in the opening ...
متن کاملMaladjustment of programmable ventricular shunt valves by inadvertent exposure to a common hospital device
Background Programmable ventricular shunt valves are commonly used to treat hydrocephalus. They can be adjusted to allow for varying amounts of cerebrospinal fluid (CSF) flow using an external magnetic programming device, and are susceptible to maladjustment from inadvertent exposure to magnetic fields. Case Description We describe the case of a 3-month-old girl treated for hydrocephalus with...
متن کاملProgrammable CSF shunt valves: radiographic identification and interpretation.
The programmable CSF shunt valve has become an important tool in hydrocephalus treatment, particularly in the NPH population and in pediatric patients with complex hydrocephalus. The purpose of this study is to provide a single reference for the identification of programmable shunt valves and the interpretation of programmable shunt valve settings. Four major manufacturers of programmable shunt...
متن کاملProgrammable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.
A programmable CSF shunt valve was assessed for magnetic field interactions, heating (transmit-receive body radio-frequency coil; whole-body averaged specific absorption rate, 2.1 W/kg), functional alterations, and artifacts at 3T. The programmable valve showed minor magnetic field interactions and heating was not excessive (+0.8 degrees C). The function of the programmable valve was not altere...