Home mechanical ventilation through mask: monitoring leakage and nocturnal oxygenation at home.

نویسندگان

  • Ramon Fernandez Alvarez
  • Gemma Rubinos Cuadrado
  • Francisco Rodriguez Jerez
  • Almudena Garcia Garcia
  • Patricia Rodriguez Menendez
  • Pere Casan Clara
چکیده

BACKGROUND Leakage is common in patients receiving home mechanical ventilation (HMV) via a face mask. Although pressure ventilators have partial compensatory capacity, excessive leakage can compromise the effectiveness of treatment. Home ventilators are equipped with built-in software which provides information on leakage. However, the values of leakage and their effects in routine clinical practice are currently little known. OBJECTIVE To measure leakage in stable patients on nocturnal HMV and its impact on treatment effectiveness. METHODS Consecutive outpatients on HMV were recruited. Nocturnal pulse oximetry was performed at home and leakage was measured using the ventilator's built-in software. We measured: mean SpO(2), percentage of time with SpO(2) <90% (T90), mean leakage (meanL), maximum leakage (maxL), and minimum leakage (minL) during the ventilation session. We estimated ventilator capacity to compensate for leakage according to inspiratory positive airway pressure and divided the patients into two groups: those with leak compensation and those without. RESULTS The study included 41 patients [mean age, 64 years (SD 11.9); 23 (56%) women]. Nocturnal pulse oximetry showed an SpO(2) of 94% (±2.9) and a T90 of 10% (±21.7). Leakage (in l/min) was: meanL, 32.2 (±15.3); maxL, 64.8 (±28.5), and minL, 18.8 (±10.6). Seven cases (17%) had leakage greater than the ventilator compensatory capacity, but no significant difference in SpO(2) or T90 was observed between patients with or without leak compensation. CONCLUSIONS A wide variation between maxL and minL was observed in our series; 17% of cases had higher leakage values than the compensatory capacity of the ventilator, but this did not affect nocturnal oxygenation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Home Mechanical Ventilation through Mask: Monitoring Leakage and Nocturnal Oxygenation at Home

Background: Leakage is common in patients receiving home mechanical ventilation (HMV) via a face mask. Although pressure ventilators have partial compensatory capacity, excessive leakage can compromise the effectiveness of treatment. Home ventilators are equipped with built-in software which provides information on leakage. However, the values of leakage and their effects in routine clinical pr...

متن کامل

Nasal versus oronasal mask in home mechanical ventilation: the preference of patients as a strategy for choosing the interface.

INTRODUCTION In home mechanical ventilation (HMV), the mask is a key factor for patient comfort and therapeutic adherence. There is no evidence on the best strategy for choosing the mask in HMV. OBJECTIVE To explore patient preference when prescribing the mask for HMV treatment and assess its relationship with effectiveness. METHODS A prospective study with repeated measures in stable pati...

متن کامل

Home mechanical ventilation

Definition and goals ofhome mechanical ventilation Born during the polio epidemics of 50 years ago, home mechanical ventilation has evolved towards two distinct concepts-a life support system for patients without respiratory independence (high cervical spinal cord injury, end stage neuromuscular disease, or chronic obstructive pulmonary disease), and elective therapy for patients with progressi...

متن کامل

Oxygen supplementation in noninvasive home mechanical ventilation: the crucial roles of CO2 exhalation systems and leakages.

BACKGROUND When supplemental oxygen is added to noninvasive ventilation using a non-ICU ventilator, it is usually introduced with a preset flow into the circuit near the ventilator; however, the impact of different CO2 exhalation systems and leaks on the actual FIO2 and gas exchange has not been elucidated. METHODS In a randomized, open-label, 4-treatment (2-by-2), 4-period crossover design, ...

متن کامل

Combined nasal intermittent positive-pressure ventilation and rocking bed in chronic respiratory insufficiency. Nocturnal ventilatory support of a disabled person at home.

The use of intermittent positive-pressure ventilation via nasal mask with a rocking bed provided the necessary ventilatory support for a person with quadriplegia living at home. This option was required to maintain an independent life-style of choice. The combination of techniques and the linkage of devices demonstrated an effective use of simple available technology for respiratory care at hom...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiration; international review of thoracic diseases

دوره 85 2  شماره 

صفحات  -

تاریخ انتشار 2013