High-field NMR using resistive and hybrid magnets.
نویسندگان
چکیده
Resistive and resistive-superconducting hybrid magnets can generate dc magnetic fields much higher than conventional superconducting NMR magnets but the field spatial homogeneity and temporal stability are usually not sufficient for high-resolution NMR experiments. Hardware and technique development addressing these issues are presented for high-resolution NMR at magnetic fields up to 40T. Passive ferromagnetic shimming and magic-angle spinning are used effectively to reduce the broadening from inhomogeneous magnetic field. A phase correction technique based on simultaneous heteronuclear detection is developed to compensate magnetic field fluctuations to achieve high spectral resolution.
منابع مشابه
High-resolution, >1 GHz NMR in unstable magnetic fields.
Resistive or hybrid magnets can achieve substantially higher fields than those available in superconducting magnets, but their spatial homogeneity and temporal stability are unacceptable for high-resolution NMR. We show that modern stabilization and shimming technology, combined with detection of intermolecular zero-quantum coherences (iZQCs), can remove almost all of the effects of inhomogenei...
متن کاملSingle Scan 2D NMR Spectroscopy on a 25 T Bitter Magnet.
2D NMR relies on monitoring systematic changes in the phases incurred by spin coherences as a function of an encoding time t(1), whose value changes over the course of independent experiments. The intrinsic multiscan nature of such protocols implies that resistive and/or hybrid magnets, capable of delivering the highest magnetic field strengths but possessing poor temporal stabilities, become u...
متن کاملBehavioral effects on rats of high strength magnetic fields generated by a resistive electromagnet.
It has been reported previously that exposure to static high magnetic fields of 7 T or above in superconducting magnets has behavioral effects on rats. In particular, magnetic field exposure acutely but transiently suppressed rearing and induced walking in tight circles; the direction of circular locomotion was dependent on the rats' orientation within the magnet. Furthermore, when magnet expos...
متن کاملHow does an MR scanner operate?
The purpose of the magnet is to provide a uniform magnetic field at all points in the bore of the magnet to align mobile protons within the patient. The strength of the magnetic field is measured in units of testa or gauss-1.0 T equals 10 000 G. The magnetic field produced by the magnet is typically 30 000 times stronger than the magnetic field of the earth. Because it is not possible to obtain...
متن کاملPermanent magnet assembly producing a strong tilted homogeneous magnetic field: towards magic angle field spinning NMR and MRI.
We introduce a cylindrical permanent magnet design that generates a homogeneous and strong magnetic field having an arbitrary inclination with respect to the axis of the cylinder. The analytical theory of 3 D magnetostatics has been applied to this problem, and a hybrid magnet structure has been designed. This structure contains two magnets producing a longitudinal and transverse component for ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of magnetic resonance
دوره 191 1 شماره
صفحات -
تاریخ انتشار 2008