Improved segmented modified Look-Locker inversion recovery T1 mapping sequence in mice
نویسندگان
چکیده
OBJECT To develop and evaluate a 2D modified Look-Locker (MOLLI) for high-resolution T1 mapping in mice using a 3T MRI scanner. MATERIALS AND METHODS To allow high-resolution T1 mapping in mice at high heart rates a multi-shot ECG-triggered 2D MOLLI sequence was developed. In the proposed T1 mapping sequence the optimal number of sampling points and pause cardiac cycles following an initial adiabatic inversion pulse was investigated in a phantom. Seven native control and eight mice, 3 days post myocardial infarction (MI) after administration of gadolinium were scanned. Two experienced readers graded the visual T1 map quality. RESULTS In T1 phantoms, there were no significant differences (<0.4% error) between 12, 15 and 20 pause cardiac cycles (p = 0.1, 0.2 and 0.6 respectively) for 8 acquisition cardiac cycles for 600bpm in comparison to the conventional inversion recovery spin echo T1 mapping sequence for short T1's (<600 ms). Subsequently, all in-vivo scans were performed with 8 data acquisitions and 12 pause cardiac cycles to minimize scan time. The mean native T1 value of myocardium in control animal was 820.5±52 ms. The post-contrast T1 measured 3 days after MI in scar was 264±59 ms and in healthy myocardium was 512±62 ms. The Bland-Altman analysis revealed mean difference of only -1.06% of infarct size percentage between T1 maps and LGE. CONCLUSIONS A multi-shot 2D MOLLI sequence has been presented that allows reliable measurement of high spatial resolution T1 maps in mice for heart rates up to 600bpm.
منابع مشابه
A segmented modified look-locker inversion recovery (MOLLI) sequence for high heart rate T1 mapping of mice
Background Quantitative T1 mapping provides myocardial tissue characterization for assessment of various cardiomyopathies. The Modified Look-locker (MOLLI) sequence is widely used for mapping the T1quantification, where multiple single-shot images are acquired along the T1 recovery curve after an inversion pulse. However, single-shot imaging becomes infeasible for mouse imaging at high heart ra...
متن کاملHow Accurate Is MOLLI T1 Mapping In Vivo? Validation by Spin Echo Methods
T1 mapping is a promising quantitative tool for assessing diffuse cardiomyopathies. The purpose of this study is to quantify in vivo accuracy of the Modified Look-Locker Inversion Recovery (MOLLI) cardiac T1 mapping sequence against the spin echo gold standard, which has not been done previously. T1 accuracy of MOLLI was determined by comparing with the gold standard inversion recovery spin ech...
متن کاملImproved precision in SASHA T1 mapping with a variable flip angle readout
Background The SAturation-recovery single-SHot Acquisition (SASHA) T1 mapping sequence has excellent accuracy independent of T1, T2, heart rate, and flip angle [1], which are known dependencies of the more commonly used MOdified Look-Locker Inversion-recovery (MOLLI) sequence. However, SASHA has a greater T1 variability (poorer precision) compared to MOLLI. A two-parameter fit, with assumed ide...
متن کاملPhantom validation of 17 and 11 heartbeat MOLLI T1 mapping sequence at 3T
Background and objective The Modified Look-Locker Inversion-Recovery (MOLLI) sequence (Messroghli, et al, JMRI, 2007) was optimized for myocardial T1 mapping on 1.5T. However, little data exists on T1 mapping at 3T. The standard MOLLI sequence uses three inversion-recovery blocks to acquire 11 images over 17 heartbeats (HB). The long breath hold could limit its clinical application in patients ...
متن کاملThe Effect of Heart Rate in Look-Locker Cardiac T1 Mapping
Introduction: Look-Locker inversion recovery produces images at multiple inversion times (TI) after a 180° inversion pulse (1). In non-cardiac and phantom experiments, the sequence timing can be carefully controlled. This allows the user to exactly specify both the TIs and the magnetization recovery time between inversions. In cardiac imaging, however, these timing variables are measured in uni...