High-performance colorimeter with an electronic bubble gate for use in miniaturized continuous-flow analyzers.

نویسندگان

  • W E Neeley
  • S C Wardlaw
  • T Yates
  • W G Hollingsworth
  • M E Swinnen
چکیده

We describe a high-performance colorimeter with an electronic bubble gate for use with miniaturized continuous-flow analyzers. The colorimeter has a flow-through cuvette with optically flat quartz windows that allows a bubbled stream to pass freely without any breakup or retention of bubbles. The fluid volume in the light path is only 1.8 mul. The electronic bubble gate selectively removes that portion of the photodector signal produced by the air bubbles passing through the flow cell and allows that portion of the signal attributable to the fluid segment to pass to the recorder. The colorimeter is easy to use, rugged, inexpensive, and requires minimal adjustments.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Design and operation of a signal comparator to increase efficiency of continuous-flow analyzers.

Design features and operation of a signal comparator are described. The efficiency of continuous-flow analyzers is improved by allowing the bubbled stream to remain intact as it passes through the colorimeter flow cell. The signal comparator eliminates the bubble artifact from the photodetector signal while allowing the signal produced from the fluid segment to remain intact. The instrument is ...

متن کامل

Peykan check, a simple continuous quality control method for hematology analyzers

A simple inexpensive con!nuous quality control method, by means of eight basic blood coun!ng parameters, obtained from automated hematology analyzers, using pa!ent samples, is described. A few samples with low, normal and high values were selected and introduced to the instrument early in the morning as the first count, the results of which are plo#ed on the appropriate chart as dots. The sa...

متن کامل

A method for detecting bubbles in two-phase gas-liquid flow

Detecting bubble in two-phase flow has been a basic issue in two-phase flow systems. A new method for measuring the frequency of bubble formation is presented in this paper. For this purpose, an electronic device was designed and constructed which works based on a change in intensity of laser beam. For this purpose, continues light beam is embedded just above the needle, which is received by a ...

متن کامل

Device for more economical standby operation of continuous-flow analyzers.

To decrease the reagent-consumption rate of continuous-flow analyzers, I constructed a device that allows standby operation. The device, when actuated by switching off the sampler, decreases the effective pumping rate to less than 1/15 normal. The bubble pattern remains virtually undisturbed, and the system can be immediately restarted for emergency use. During five months of use in our laborat...

متن کامل

Performance Investigation of Pentacene Based Organic Double Gate Field Effect Transistor and its Application as an Ultrasensitive Biosensor

In this paper, the electrical performance of double gate organic field effecttransistor (DG-OFET) are thoroughly investigated and feasibility of the deviceas an efficient biosensor is comprehensively assessed. The introduced deviceprovides better gate control over the channel, yielding better charge injectionproperties from source to channel and providing higher on-state...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 22 2  شماره 

صفحات  -

تاریخ انتشار 1976