Comparison of the BACTEC MGIT 960 and BACTEC 460TB systems for detection of mycobacteria in clinical specimens.
نویسندگان
چکیده
The reliability of the Mycobacteria Growth Indicator Tube (MGIT) 960 system for rapid detection of mycobacteria in clinical specimens was evaluated and compared to the radiometric method (BACTEC 460TB) and to mycobacterial culture on Lowenstein-Jensen (LJ) medium. Clinical specimens (n = 590) were tested without selection. A total of 121 (20.5%) isolates of mycobacteria were recovered; 98 (81.0%) of them were recovered with the BACTEC 460TB system, 86 (71.1%) were recovered with the BACTEC MGIT 960 system, and 55 (45.5%) were recovered with LJ medium (MGIT 960 versus BACTEC 640TB, p >0.05; MGIT 960 or BACTEC 460TB versus LJ, p <0.001). The mean time to detection (TTD) was 18 da for BACTEC 460 TB, and 13 da for BACTEC MGIT 960. The mean time to detection in each system, based upon data where both systems were culture positive, was significantly different (16.6 da for BACTEC 460TB and 13 da for BACTEC MGIT 960, p<0.001). The contamination rate of the BACTEC MGIT 960 system was 13.2%, which was intermediate between the BACTEC 460TB system (11.7%) and the LJ medium (14.7%). These data indicate that the fully automated MGIT 960 system is an accurate, non-radiometric alternative to the BACTEC 460TB method for rapid detection of mycobacteria in a clinical setting.
منابع مشابه
Evaluation of BACTEC MGIT 960 and BACTEC 460TB systems for recovery of mycobacteria from clinical specimens of a university hospital with low incidence of tuberculosis.
Clinical samples obtained over a period of 8 months (n = 2,624) were processed in parallel with the BACTEC 460TB system, with the MGIT 960 system, and in Löwenstein-Jensen (LJ) medium, resulting in the recovery of 127 mycobacteria. Recovery rates in combinations of the BACTEC 460TB or MGIT 960 system with LJ were, respectively, 94.7 and 94.7% for Mycobacterium tuberculosis complex (n = 57) and ...
متن کاملComparison of BACTEC MGIT 960 system and BACTEC 460 TB system for growth and detection of Mycobacteria from clinical specimens.
This study was carried out to compare the performance of BACTEC MGIT 960 with the BACTEC 460 TB for growth and detection of Mycobacteria from human clinical specimens. The BACTEC MGIT 960 instrument is a fully automated system that utilizes MGIT tubes containing an oxygen sensor embedded in silicon at the bottom and filled with 7 mL of modified Middlebrook 7H9 broth. Identical samples were inoc...
متن کاملMulticenter laboratory evaluation of the MB/BacT Mycobacterium detection system and the BACTEC MGIT 960 system in comparison with the BACTEC 460TB system for susceptibility testing of Mycobacterium tuberculosis.
In this multicenter study, the reliability of two nonradiometric, fully automated systems, the MB/BacT and BACTEC MGIT 960 systems, for testing the susceptibilities of 82 Mycobacterium tuberculosis strains to isoniazid, rifampin, ethambutol, and streptomycin was evaluated in comparison with the radiometric BACTEC 460TB system. The arbitration of discrepant results was done by the reanalysis of ...
متن کاملComparison of the BACTEC MGIT 960 and ESP culture system II for growth and detection of mycobacteria.
The performances of two continuously monitoring mycobacterial culture systems-ESP Culture System II (ESP II; Trek Diagnostics, Inc. , Westlake, Ohio) and BACTEC MGIT 960 (BD Biosciences, Sparks, Md. )-were compared. In addition to both liquid media, all specimens were plated onto Middlebrook 7H11/7H11 selective agar. A total of 3, 151 specimens of all types (56.3% were respiratory specimens) we...
متن کاملMulticenter evaluation of the BACTEC MGIT 960 system for recovery of mycobacteria.
We evaluated the BACTEC MGIT 960 system, which is a fully automated, noninvasive system for the growth and detection of mycobacteria with a capacity to incubate and continuously monitor 960 7-ml culture tubes. We studied 3,330 specimens, 2,210 respiratory and 1,120 nonrespiratory specimens, collected from 2,346 patients treated at six sites. Processed specimens were inoculated into the BACTEC M...