Calibration Test of PET Scanners in a Multi-Centre Clinical Trial on Breast Cancer Therapy Monitoring Using 18F-FLT

نویسندگان

  • Francis Bouchet
  • Lilli Geworski
  • Bernd O. Knoop
  • Ludovic Ferrer
  • Alina Barriolo-Riedinger
  • Corinne Millardet
  • Marjolaine Fourcade
  • Antoine Martineau
  • Anne Belly-Poinsignon
  • Francis Djoumessi
  • Karine Tendero
  • Laurent Keros
  • Frederic Montoya
  • Christel Mesleard
  • Anne-Laure Martin
  • Franck Lacoeuille
  • Olivier Couturier
چکیده

UNLABELLED A multi-centre trial using PET requires the analysis of images acquired on different systems We designed a multi-centre trial to estimate the value of 18F-FLT-PET to predict response to neoadjuvant chemotherapy in patients with newly diagnosed breast cancer. A calibration check of each PET-CT and of its peripheral devices was performed to evaluate the reliability of the results. MATERIAL AND METHODS 11 centres were investigated. Dose calibrators were assessed by repeated measurements of a 68Ge certified source. The differences between the clocks associated with the dose calibrators and inherent to the PET systems were registered. The calibration of PET-CT was assessed with an homogeneous cylindrical phantom by comparing the activities per unit of volume calculated from the dose calibrator measurements with that measured on 15 Regions of Interest (ROIs) drawn on 15 consecutive slices of reconstructed filtered back-projection (FBP) images. Both repeatability of activity concentration based upon the 15 ROIs (ANOVA-test) and its accuracy were evaluated. RESULTS There was no significant difference for dose calibrator measurements (median of difference -0.04%; min = -4.65%; max = +5.63%). Mismatches between the clocks were less than 2 min in all sites and thus did not require any correction, regarding the half life of 18F. For all the PET systems, ANOVA revealed no significant difference between the activity concentrations estimated from the 15 ROIs (median of difference -0.69%; min = -9.97%; max = +9.60%). CONCLUSION No major difference between the 11 centres with respect to calibration and cross-calibration was observed. The reliability of our 18F-FLT multi-centre clinical trial was therefore confirmed from the physical point of view. This type of procedure may be useful for any clinical trial involving different PET systems.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In vivo imaging of cellular proliferation in renal cell carcinoma using 18F-fluorothymidine PET

Objective(s): The ability to measure cellular proliferation non-invasively in renal cell carcinoma may allow prediction of tumour aggressiveness and response to therapy. The aim of this study was to evaluate the uptake of 18Ffluorothymidine (FLT) PET in renal cell carcinoma (RCC), and to compare this to 18F-fluorodeoxyglucose (FDG), and to an immunohistochemical measure of cellular proliferatio...

متن کامل

Imaging biomarkers predict response to anti-HER2 (ErbB2) therapy in preclinical models of breast cancer.

PURPOSE To evaluate noninvasive imaging methods as predictive biomarkers of response to trastuzumab in mouse models of HER2-overexpressing breast cancer. The correlation between tumor regression and molecular imaging of apoptosis, glucose metabolism, and cellular proliferation was evaluated longitudinally in responding and nonresponding tumor-bearing cohorts. EXPERIMENTAL DESIGN Mammary tumor...

متن کامل

Triage of Limited Versus Extensive Disease on 18F-FDG PET/CT Scan in Small Cell lung Cancer

Objective(s): Small cell lung cancer (SCLC) is an aggressive neuroendocrine carcinoma, which accounts for 10-15% of pulmonary cancers and exhibits early metastatic spread. This study aimed to determine the added value of 18F-FDG PET/CT imaging in tumor, node, and metastasis (TNM) staging of SCLC, compared to the conventional computed tomography (CT) scan and its potential role as a prognosticat...

متن کامل

Imaging, Diagnosis, Prognosis [18F]-30Deoxy-30-Fluorothymidine Positron Emission Tomography and Breast Cancer Response to Docetaxel

Purpose: To establish biomarkers indicating clinical response to taxanes, we determined whether early changes in [18F]-30deoxy-30-fluorothymidine positron emission tomography (FLT-PET) can predict benefit from docetaxel therapy in breast cancer. Experimental Design: This was a prospective unblinded study in 20 patients with American Joint Committee on Cancer (AJCC) stage II–IV breast cancer unr...

متن کامل

Clinical Impact of 18F-FDG PET/CT on the Management of Gynecologic Cancers: One Center Experience

Objective(s): We aim to investigate the clinical impact of 18F-FDG PET/CT in managing patients with gynecological malignancies and pelvic or extrapelvic lymph nodes that are of equivocal significance on conventional imaging.Methods: We retrospectively evaluated 18F-FDG PET/CT scans of patients with gynecologic tumors who were referred to King Hussein Cancer Center from January 2010 to August 20...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2013