Evaluation of the LIAISON chemiluminescence immunoassay for diagnosis of syphilis.
نویسندگان
چکیده
We report the results of an evaluation of the LIAISON Treponema pallidum-specific assay, a one-step sandwich chemiluminescence immunoassay (CLIA), as a screening test and as a confirmatory test for the diagnosis of syphilis. The assay was compared with the CAPTIA Syphilis-G enzyme immunoassay (EIA) and with a testing algorithm that also included the rapid plasma reagin (RPR) and T. pallidum particle agglutination (PA) assays. As a screening test, the CLIA showed levels of agreement with the EIA and with the algorithm, respectively, of 94.1 and 100% for 51 samples from patients with primary or secondary syphilis, 93.2 and 98.7% for 999 samples sent to the laboratory for routine syphilis testing, 84.5 and 94.0% for 200 samples from human immunodeficiency virus-positive patients, 98.0 and 100% for 200 samples from pregnant patients, and 94.3 and 98.3% for 992 samples from apparently healthy adults. As a confirmatory test, the CLIA showed 99% agreement with the EIA for 204 RPR-positive samples. After resolution with further T. pallidum PA testing and the discarding of one sample of insufficient quantity, there was 100% agreement for the remaining 203 samples. For the total group of 2,645 samples, the overall relative sensitivity was 95.8% and the relative specificity was 99.1%. We conclude that the LIAISON CLIA demonstrated excellent sensitivity and specificity when evaluated as a confirmatory test and as a screening test for syphilis among various patient populations, including specific populations with reportedly increased rates of false-positive nontreponemal test results.
منابع مشابه
Evaluation of LIAISON Treponema Screen, a novel recombinant antigen-based chemiluminescence immunoassay for laboratory diagnosis of syphilis.
The purpose of this study was to evaluate the diagnostic performance of LIAISON Treponema Screen (DiaSorin, Saluggia, Italy), a new automated chemiluminescence immunoassay (CLIA), in comparison with that of rapid plasma reagin (RPR) and the following currently used treponemal tests: hemagglutination test (TPHA), immunoenzymatic assay (EIA), and Western blot (WB). First, a retrospective study wa...
متن کاملEvaluation of a new chemiluminescence immunoassay for diagnosis of syphilis
OBJECTIVE To assess the sensitivity, specificity, and feasibility of a new chemiluminescence immunoassay (CLIA) in the diagnosis of syphilis. METHODS At first, a retrospective study was conducted, using 135 documented cases of syphilis and 30 potentially interfering samples and 80 normal sera. A prospective study was also performed by testing 2, 071 unselected samples for routine screening fo...
متن کاملEvaluation of the Vitros Syphilis TPA chemiluminescence immunoassay as a first-line method for reverse syphilis screening.
We report here the results of the diagnostic performances of Vitros Syphilis TPA (a chemiluminescence treponemal assay) compared with those of two treponemal enzyme immunoassays and of traditional versus reverse syphilis algorithms. Ease of use, automation, and high throughput make the Vitros Syphilis TPA assay a good choice for syphilis screening in high-volume laboratories.
متن کاملSerologic diagnosis of syphilis: comparison of different diagnostic methods.
Diagnostic and therapeutic approaches in syphilis show wide variation. The use of only one type of serologic test is insufficient for diagnosis. However, current international recommendations cannot be applied due to various reasons (cost, availability, etc.). The aim of the study was to review serologic data of syphilis patients to determine diagnostic performance of three different methods. I...
متن کاملSerological diagnosis of syphilis: a comparison of different diagnostic methods.
INTRODUCTION Serological tests' limitations in syphilis diagnosis as well as numerous test interpretations mean that patients with discordant serology results can present diagnostic and treatment challenges for clinicians. We analyzed three common diagnostic algorithms for detecting suspected syphilis in high-prevalence populations in Slovenia. METHODS The prospective study included a total o...