Planimetry of aortic valve area using multiplane transoesophageal echocardiography is not a reliable method for assessing severity of aortic stenosis.

نویسندگان

  • Y Bernard
  • N Meneveau
  • A Vuillemenot
  • D Magnin
  • T Anguenot
  • F Schiele
  • J P Bassand
چکیده

OBJECTIVE To assess the reliability of aortic valve area planimetry by multiplane transoesophageal echocardiography (TOE) in aortic stenosis. DESIGN Study of the diagnostic value of aortic valve area planimetry using multiplane TOE, compared with catheterisation and the continuity equation, both being considered as criterion standards. SETTING University hospital. PATIENTS 49 consecutive patients (29 male, 20 female, aged 44 to 82 years, average 66.6 (SD 8.5)), referred for haemodynamic evaluation of an aortic stenosis, were enrolled in a prospective study. From this sample, 37 patients were eligible for the final analysis. METHODS Transthoracic and multiplane transoesophageal echocardiograms were performed within 24 hours before catheterisation. At transthoracic echo, aortic valve area was calculated by the continuity equation. At TOE, the image of the aortic valve opening was obtained with a 30-65 degrees rotation of the transducer. Numerical dynamic images were stored on optical discs for off-line analysis and were reviewed by two blinded observers. Catheterisation was performed in all cases and aortic valve area was calculated by the Gorlin formula. RESULTS Feasibility of the method was 92% (48/52). The agreement between aortic valve area measured at TOE (mean 0.88 (SD 0.35) cm2) and at catheterisation (0.79 (0.24) cm2) was very poor. The same discrepancies were found between TOE and the continuity equation (0.72 (0.26) cm2). TOE planimetry overestimated aortic valve area determined by the two other methods. Predictive positive and negative values of planimetry to detect aortic valve area < 0.75 cm2 were 62% (10/16) and 43% (9/21) respectively. CONCLUSIONS Planimetry of aortic valve area by TOE is difficult and less accurate than the continuity equation for assessing the severity of aortic stenosis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

CARDIOVASCULAR MEDICINE Evaluation of aortic stenosis by cardiovascular magnetic resonance imaging: comparison with established routine clinical techniques

Objective: To evaluate whether direct planimetry of aortic valve area (AVA) by cardiac magnetic resonance (CMR) imaging is a reliable tool for determining the severity of aortic stenosis compared with transthoracic echocardiography (TTE), transoesophageal echocardiography (TOE), and cardiac catheterisation. Methods: 44 symptomatic patients with severe aortic stenosis were studied. By cardiac ca...

متن کامل

Evaluation of aortic stenosis by cardiovascular magnetic resonance imaging: comparison with established routine clinical techniques.

OBJECTIVE To evaluate whether direct planimetry of aortic valve area (AVA) by cardiac magnetic resonance (CMR) imaging is a reliable tool for determining the severity of aortic stenosis compared with transthoracic echocardiography (TTE), transoesophageal echocardiography (TOE), and cardiac catheterisation. METHODS 44 symptomatic patients with severe aortic stenosis were studied. By cardiac ca...

متن کامل

What do you mean by aortic valve area: geometric orifice area, effective orifice area, or gorlin area?

Aortic valve area can be measured by cardiac catheterization, Doppler echocardiography, or imaging planimetry to assess aortic stenosis severity. These diagnostic techniques provide the Gorlin area, the effective orifice area (EOA) and the geometric orifice area (GOA), respectively. The differences between these three parameters depend mainly on the valve inflow shape and cross-sectional area o...

متن کامل

Severe aortic valve stenosis with low-gradient and preserved ejection fraction: a misclassification issue?

INTRODUCTION AND OBJECTIVES Low-gradient severe aortic stenosis with preserved ejection fraction is a controversial entity. Misclassification of valvulopathy severity could explain the inconsistencies reported in the prognosis of these patients. Planimetry of the aortic area using three-dimensional transesophageal echocardiography could clear up these doubts. The objectives were to assess the a...

متن کامل

Magnetic resonance to assess the aortic valve area in aortic stenosis: how does it compare to current diagnostic standards?

OBJECTIVES The purpose of the present study was to evaluate whether magnetic resonance (MR) planimetry of the aortic valve area (AVA) may prove to be a reliable, non-invasive diagnostic tool in the assessment of aortic valve stenosis, and how the results compare with current diagnostic standards. BACKGROUND Current standard techniques for assessing the severity of aortic stenosis include tran...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Heart

دوره 78 1  شماره 

صفحات  -

تاریخ انتشار 1997