Can endotracheal bioimpedance cardiography assess hemodynamic response to passive leg raising following cardiac surgery?
نویسندگان
چکیده
BACKGROUND The utility of endotracheal bioimpedance cardiography (ECOM) has been scarcely reported. We tested the hypothesis that it could be an alternative to pulse contour analysis for cardiac index measurement and prediction in fluid responsiveness. METHODS Twenty-five consecutive adult patients admitted to the intensive care unit following conventional cardiac surgery were prospectively included and investigated at baseline, during passive leg raising, and after fluid challenge. Comparative cardiac index data points were collected from pulse contour analysis (CIPC) and ECOM (CIECOM). Correlations were determined by linear regression. Bland-Altman analysis was used to compare the bias, precision, and limits of agreement. Percentage error was calculated. Changes in CIPC (ΔCIPC) and CIECOM (ΔCIECOM) during passive leg raising were collected to assess their discrimination in predicting fluid responsiveness. RESULTS A significant relationship was found between CIPC and CIECOM (r = 0.45; P < 0.001). Bias, precision, and limits of agreement were 0.44 L.min-1.m-2 (95% confidence interval, 0.33-0.56), 0.59 L.min-1.m-2, and -0.73 to 1.62 L.min-1.m-2, respectively. Percentage error was 45%. A significant relationship was found between percent changes in CIPC and CIECOM after fluid challenge (r = 0.42; P = 0.035). Areas under the ROC curves for ΔCIPC and ΔCIECOM to predict fluid responsiveness were 0.72 (95% confidence interval, 0.5-0.88) and 0.81 (95% confidence interval, 0.61-0.94), respectively. CONCLUSIONS ECOM is not interchangeable with pulse contour analysis but seems consistent to monitor cardiac index continuously and could help to predict fluid responsiveness by using passive leg raising.
منابع مشابه
Effect of Passive Leg Raising on Systemic Hemodynamics of Pregnant Women: A Dynamic Assessment of Maternal Cardiovascular Function at 22–24 Weeks of Gestation
OBJECTIVE To investigate functional hemodynamic response to passive leg raising in healthy pregnant women and compare it with non-pregnant controls. MATERIALS AND METHODS This was a prospective cross-sectional study with a case-control design. A total of 108 healthy pregnant women at 22-24 weeks of gestation and 54 non-pregnant women were included. Cardiac function and systemic hemodynamics w...
متن کاملValues of hemodynamic variation in response to passive leg raising in predicting exercise capacity of heart failure with preserved ejection fraction
In heart failure patients with preserved ejection fraction, their hemodynamic parameters usually change when they are from recumbent to passive leg raising. The authors designed this study to investigate the relationship between hemodynamic parameters measured by impedance cardiography (ICG) and 6-minute walk distance (6MWD) of heart failure with preserved ejection fraction (HFPEF). We recruite...
متن کاملBioimpedance cardiography measurements of cardiac output and other cardiovascular parameters.
This paper describes impedance technique and device types, discusses hemodynamic data parameters that are available, discusses the differences between impedance cardiography and data that are derived from invasive pulmonary artery catheters, and explains how nurses can apply bioimpedance cardiography in a variety of patient populations.
متن کاملStatic and Functional Hemodynamic Profiles of Women with Abnormal Uterine Artery Doppler at 22–24 Weeks of Gestation
OBJECTIVE To compare cardiac function, systemic hemodynamics and preload reserve of women with increased (cases) and normal (controls) uterine artery (UtA) pulsatility index (PI) at 22-24 weeks of gestation. MATERIALS AND METHODS A prospective cross-sectional study of 620 pregnant women. UtA blood flow velocities were measured using Doppler ultrasonography, and PI was calculated. Mean UtA PI ...
متن کاملMaternal Functional Hemodynamics in the Second Half of Pregnancy: A Longitudinal Study
OBJECTIVE Cardiovascular response to passive leg raising (PLR) is useful in assessing preload reserve, but it has not been studied longitudinally during pregnancy. We aimed to investigate gestational age associated serial changes in maternal functional hemodynamics and establish longitudinal reference ranges for the second half of pregnancy. MATERIALS AND METHODS This was a prospective longit...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 2 شماره
صفحات -
تاریخ انتشار 2012