Feasibility of Macular Integrity Assessment (MAIA) Microperimetry in Children: Sensitivity, Reliability, and Fixation Stability in Healthy Observers.

نویسندگان

  • Pete R Jones
  • Narmella Yasoubi
  • Marko Nardini
  • Gary S Rubin
چکیده

Purpose To assess the feasibility of macular integrity assessment (MAIA) microperimetry (MP) in children. Also to establish representative outcome measures (differential light sensitivity, fixation stability, test-retest reliability) for children without visual impairment. Methods Thirty-three adults and 33 children (9-12 years) were asked to perform three monocular MAIA examinations within a single session (dominant eye only). Results Children exhibited poorer test-retest reliability than adults for measures of both mean sensitivity (95% coefficient of repeatability [CoR95] = 2.7 vs. 2.3 dB, P = 0.036) and pointwise sensitivity (CoR95 = 6.2 vs. 5.7 dB, P < 0.001). Mean sensitivity was lower in children (27.6 vs. 29.5 dB, P < 0.001), and fixation stability was poorer (95% bivariate contour ellipse area [BCEA95] = 4.58 vs. 1.14, P < 0.001). Mean sensitivity was negatively correlated with fixation stability (r = -0.44, P < 0.001). Both children and adults exhibited substantial practice effects, with mean sensitivity improving by 0.5 dB (adults) and 0.9 dB (children) between examinations 1 and 2 (P ≤ 0.017). There were no significant differences between examinations 2 and 3 (P ≥ 0.374). Conclusions Microperimetry is feasible in 9- to 12-year-old children. However, systematically lower sensitivities mean that the classification boundary for "healthy" performance should be lowered in children, pending development of techniques to improve attentiveness/fixation that may reduce or remove this difference. High measurement variability suggests that the results of multiple tests should be averaged when possible. Learning effects are a potential confound, and it is recommended that the results of the first examination be discarded.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Visual rehabilitation via microperimetry in patients with geographic atrophy: a pilot study

BACKGROUND Age-related macular degeneration (AMD) is the leading cause of blindness in the western world. As a consequence of AMD, patients develop structural damage that comprises the fovea and subsequently present loss of central vision, low visual acuity and unstable fixation. Contrary to what happens with anti-angiogenic treatment in neovascular AMD, there is currently no definitive treatme...

متن کامل

Correlation Between Macular Integrity Assessment and Optical Coherence Tomography Imaging of Ellipsoid Zone in Macular Telangiectasia Type 2.

Purpose To correlate ellipsoid zone (EZ) defects on spectral-domain optical coherence tomography (SD-OCT) with retinal sensitivity loss on macular integrity assessment (MAIA) microperimetry in macular telangiectasia type 2 (MacTel). Methods Macular SD-OCT volumes and microperimetry maps were obtained during the international, multicenter, randomized phase 2 trial of ciliary neurotrophic facto...

متن کامل

Microperimetric Biofeedback Training Improved Visual Acuity after Successful Macular Hole Surgery.

Purpose. To evaluate the efficacy of setting a preferred retinal locus relocation target (PRT) and performing Macular Integrity Assessment (MAIA) biofeedback training in patients showing insufficient recovery of best corrected visual acuity (BCVA) despite successful closure of an idiopathic macular hole (MH). Methods. Retrospective interventional case series. Nine eyes of 9 consecutive patients...

متن کامل

Macular sensitivity in eyes with central serous chorioretinopathy.

PURPOSE To determine macular sensitivity and fixation characteristics in eyes with central serous chorioretinopathy (CSC) using fundus-related microperimetry. METHODS The authors reviewed 19 eyes with serous elevation within the central 10 degrees due to CSC and 15 normal healthy eyes that had undergone fundus-related microperimetry. The macular sensitivity was measured using the fundus-relat...

متن کامل

Effective Dynamic Range and Retest Reliability of Dark-Adapted Two-Color Fundus-Controlled Perimetry in Patients With Macular Diseases.

Purpose To determine the effective dynamic range (EDR), retest reliability, and number of discriminable steps (DS) for mesopic and dark-adapted two-color fundus-controlled perimetry (FCP) using the S-MAIA (Scotopic-Macular Integrity Assessment) "micro-perimeter." Methods In this prospective cross-sectional study, each of the 52 eyes of 52 subjects with various macular diseases (mean age 62.0 ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Investigative ophthalmology & visual science

دوره 57 14  شماره 

صفحات  -

تاریخ انتشار 2016