Discriminating between normal and glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical Coherence Tomograph.
نویسندگان
چکیده
OBJECTIVE To compare the ability of 3 instruments, the Heidelberg Retina Tomograph (HRT), the GDx Nerve Fiber Analyzer (GDx), and the Optical Coherence Tomograph (OCT), to discriminate between healthy eyes and eyes with early to moderate glaucomatous visual field loss. SUBJECTS AND METHODS Forty-one patients with early to moderate glaucomatous visual field loss and 50 healthy subjects were included in the study. The HRT, GDx, and OCT imaging and visual field testing were completed on 1 eye from each subject within a 6-month interval. Statistical differences in sensitivity at fixed specificities of 85%, 90%, and 95% were evaluated. In addition, areas under the receiver operating characteristic (ROC) curve were compared. RESULTS No significant differences were found between the area under the ROC curve and the best parameter from each instrument: OCT thickness at the 5-o'clock inferior temporal position (mean +/- SE, 0.87 +/- 0.04), HRT mean height contour in the nasal inferior region (mean +/- SE, 0.86 +/- 0.04), and GDx linear discriminant function (mean +/- SE, 0.84 +/- 0.04). Twelve HRT, 2 GDx, and 9 OCT parameters had an area under the ROC curve of at least 0.81. At a fixed specificity of 90%, significant differences were found between the sensitivity of OCT thickness at the 5-o'clock inferior temporal position (71%) and parameters with sensitivities less than 52%. Qualitative assessment of stereophotographs resulted in a sensitivity of 80%. CONCLUSION Although the area under the ROC curves was similar among the best parameters from each instrument, qualitative assessment of stereophotographs and measurements from the OCT and HRT generally had higher sensitivities than measurements from the GDx.
منابع مشابه
Using optical imaging summary data to detect glaucoma.
PURPOSE To evaluate the sensitivity and specificity for discriminating between early to moderate glaucomatous and normal eyes using summary data reports from the Heidelberg Retina Tomograph (HRT), the GDx Nerve Fiber Analyzer (GDx), and the Optical Coherence Tomograph (OCT). DESIGN Comparative cross-sectional study PARTICIPANTS One eye each of 50 normal subjects and 39 glaucoma patients wit...
متن کاملRetinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer.
PURPOSE To compare retinal nerve fiber layer (RNFL) thickness measured by Heidelberg retina tomograph (HRT) and nerve fiber analyzer (GDx). METHODS Twenty eyes of 20 consecutive healthy subjects were recruited for this study. Each subject had a normal visual field and a normal optic nerve head, which was assessed by slit-lamp biomicroscopy using a 90 degrees lens. Using the HRT and GDx, RNFL ...
متن کامل[Relationship between standard automated perimetry and HRT, OCT and GDx in normal, ocular hypertensive and glaucomatous subjects].
PURPOSE To determine the correlations between structural parameters of the optic nerve head and the retinal nerve fiber layer (RNFL) obtained by using a Heidelberg Retina Tomograph II (HRT II), optic coherence tomography (OCT), and laser polarimetry (GDX-VCC) and the perimetric indices of standard automated perimetry (SAP) in normal, ocular hypertensive and glaucomatous subjects. METHODS Thre...
متن کاملDiffuse glaucomatous structural and functional damage in the hemifield without significant pattern loss.
OBJECTIVES To compare the retinal nerve fiber layer (RNFL) thickness and retinal sensitivity in the normal visual hemifield of glaucomatous eyes with localized visual field loss with those of normal eyes and eyes with suspected glaucoma, and to evaluate the relationship between RNFL atrophy and glaucoma severity. METHODS One randomly selected eye of each subject underwent standard automated p...
متن کاملComparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomograph for the detection of glaucoma.
OBJECTIVE To compare the abilities of current commercially available versions of 3 optical imaging techniques: scanning laser polarimetry with variable corneal compensation (GDx VCC), confocal scanning laser ophthalmoscopy (HRT II [Heidelberg Retina Tomograph]), and optical coherence tomography (Stratus OCT) to discriminate between healthy eyes and eyes with glaucomatous visual field loss. ME...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Archives of ophthalmology
دوره 119 7 شماره