Retrieval analysis of metal and ceramic femoral heads on a single CoCr stem design
نویسندگان
چکیده
OBJECTIVES The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design. METHODS This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40 trunnions paired with either CoCr (n = 21) or ceramic (n = 27) heads. The taper junction of all hips was evaluated for fretting/corrosion damage and volumetric material loss using a roundness-measuring machine. We used linear regression analysis to investigate taper damage differences after adjusting for potential confounding variables. RESULTS We measured median taper material loss rates of 0.210 mm3/year (0.030 to 0.448) for the metal head group and 0.084 mm3/year (0.059 to 0.108) for the ceramic group. The difference was not significant (p = 0.58). Moreover, no significant correlation between material loss and implant or patient factors (p > 0.05) was found. CONCLUSIONS Metal heads did not increase taper damage on CoCr trunnions compared with ceramic heads from the same hip design. The amount of material released at the taper junctions was very low when compared with available data regarding CoCr/Ti coupling in metal-on-metal bearings.Cite this article: A. Di Laura, H. Hothi, J. Henckel, I. Swiatkowska, M. H. L. Liow, Y-M. Kwon, J. A. Skinner, A. J. Hart. Retrieval analysis of metal and ceramic femoral heads on a single CoCr stem design. Bone Joint Res 2017;6:-350. DOI: 10.1302/2046-3758.65.BJR-2016-0325.R1.
منابع مشابه
Do Ceramic Femoral Heads Reduce Taper Fretting Corrosion in Hip Arthroplasty? A Retrieval Study
BACKGROUND Previous studies regarding modular head-neck taper corrosion were largely based on cobalt chrome (CoCr) alloy femoral heads. Less is known about head-neck taper corrosion with ceramic femoral heads. QUESTIONS/PURPOSES We asked (1) whether ceramic heads resulted in less taper corrosion than CoCr heads; (2) what device and patient factors influence taper fretting corrosion; and (3) w...
متن کاملDoes Metal Transfer Differ on Retrieved Ceramic and CoCr Femoral Heads?
Metal transfer has been observed on retrieved THA femoral heads for both CoCr and ceramic bearing materials. In vitro wear testing has shown increased wear to polyethylene acetabular liners with the presence of metal transfer. This study sought to investigate the extent of metal transfer on the bearing surface of CoCr and ceramic femoral heads and identify prevalent morphologies. Three bearing ...
متن کاملFretting and Corrosion Damage at the Head-Neck Taper is Reduced with Ceramic Femoral Heads: A Retrieval Study
Introduction: Modular head-neck total hip arthroplasty (THA) offers several benefits for orthopaedic surgeons, including intraoperative flexibility and the ability to change the head at revision surgery [1]. However, modularity at the head-neck taper, especially the use of metallic heads and stems, can lead to fretting and corrosion at the taper junction, leading to metal ions and debris releas...
متن کاملSurface Roughness of CoCr and ZrO2 Femoral Heads with Metal Transfer: A Retrieval and Wear Simulator Study
Metal transfer to femoral heads may result from impingement against the metallic acetabular shell following subluxation/dislocation, or when metallic debris enters the articulation zone. Such transfers roughen the head surface, increasing polyethylene wear in total hip replacements. Presently, we examined the surface roughness of retrieved femoral heads with metallic transfer. Profilometry reve...
متن کاملLong-term survival of a cemented titanium-aluminium-vanadium alloy straight-stem femoral component.
We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina ceramic (50 heads). Of the study group, 55 patients (55 stems) had died and six (six stems) were lost to follow-up. At a mean of 13.1 years (3 to 15.3) 26 stems had been revised for aseptic loosening. The mean f...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 6 شماره
صفحات -
تاریخ انتشار 2017