Carotid stenting using heparin-coated balloon-expandable stent reduces intimal hyperplasia in a baboon model.
نویسندگان
چکیده
PURPOSE This study evaluates the effect of heparin-coated balloon-expandable stents on intimal hyperplasia following carotid artery stenting in a baboon model. MATERIALS AND METHODS Balloon-expandable (Palmaz-Schatz) stents were placed in bilateral common carotid arteries in 26 male baboons (mean weight: 11.3 kg). In each animal, a heparin-coated (HC) carotid stent was placed on one side, whereas the contralateral carotid artery received an uncoated stent that served as a control. The carotid stents were harvested at 30 days (n = 13) and 90 days (n = 13). Arteriography was performed to assess the carotid patency, and intravascular ultrasound was used to determine neointimal and luminal areas. Histological, morphometric analysis, and scanning electron microscopy were performed in the stented carotid arteries. RESULTS One animal was excluded in each of the 1-month and 3-month groups because of premature death. Ten control (83%) and 11 HC stents (92%) remained patent in the 30-day group. In contrast, 10 control (83%) and 10 HC stents (83%) remained patent in the 90-day group. Morphometric analysis of the 30-day group showed the HC-stented carotid arteries had larger luminal areas (13%, P < 0.05), less neointimal areas (38%, P < 0.05), less neointimal/media ratios (53%, P < 0.05), and equivalent medial areas (NS) when compared with the control group. Morphometric analysis of the 90-day data showed the HC-stented carotid arteries had less neointimal areas (38%, P < 0.05) and less neointimal/media ratios (48%, P < 0.05), whereas the luminal areas and medial areas remained equivalent (NS) when compared with the control group. CONCLUSIONS Carotid stenting using heparin-coated stents reduces early intimal hyperplasia in a baboon model. This approach may represent a useful strategy for improving luminal patency in endovascular carotid intervention.
منابع مشابه
Self-expanding and balloon-expandable stents in the treatment of carotid aneurysms: an experimental study in a canine model.
PURPOSE To assess the efficacy of metal stents for the treatment of different forms and sizes of carotid aneurysms. METHODS A total of 14 experimentally constructed aneurysms in dogs were treated with transfemorally placed balloon-expandable tantalum and self-expanding nitinol stents. RESULTS In 10 cases, stenting produced either immediate complete occlusion of the aneurysm (n = 7) or compl...
متن کاملIncreased intimal hyperplasia in experimental vein graft stenting compared to arterial stenting: comparisons in a new rabbit model of stent injury.
BACKGROUND In-stent restenosis due to intimal hyperplasia is an important clinical problem. Animal models of stent injury are limited by inconsistent arterial responses to stenting, and less intimal hyperplasia than diseased human vessels. To address these issues, we aimed to compare the degree of intimal hyperplasia in stented rabbit jugular-carotid interposition grafts (vein grafts) versus st...
متن کاملInterleukin-10 inhibits intimal hyperplasia after angioplasty or stent implantation in hypercholesterolemic rabbits.
BACKGROUND Intimal hyperplasia after stent implantation is the main cause of in-stent restenosis. Activated monocytes play a key role in intimal growth. The anti-inflammatory cytokine interleukin-10 (IL-10) is a potent monocyte deactivator, endogenously produced in the atherosclerotic plaque. We tested the hypothesis that exogenous IL-10 may limit postangioplasty intimal hyperplasia after ballo...
متن کاملTwo Cases of Carotid Artery Stenting Combined Balloon- and Self-expanding Stent for the Spontaneous Internal Carotid Artery Dissections
Objective: Although conservative treatment is the first choice for high cervical internal carotid artery (ICA) stenosis due to spontaneous cervical ICA dissection, surgical revascularization is necessary when the disease resists conservative treatment. Although such lesions are difficult to treat by conventional carotid endarterectomy (CEA) or carotid artery stenting (CAS), they may be treated ...
متن کاملIncreased in-stent stenosis in ApoE knockout mice: insights from a novel mouse model of balloon angioplasty and stenting.
OBJECTIVE We aimed to develop and validate a model of angioplasty and stenting in mice that would allow investigation of the response to stent injury using genetically modified mouse strains. METHODS AND RESULTS Aortic segments from either C57BL/6 wild-type or atherosclerotic ApoE-KO mice underwent balloon angioplasty alone or balloon angioplasty and stenting with a 1.25x2.5 mm stainless stee...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of surgical research
دوره 112 1 شماره
صفحات -
تاریخ انتشار 2003