Repetitive optimizing left ventricular pacing configurations with quadripolar leads improves response to cardiac resynchronization therapy
نویسندگان
چکیده
BACKGROUND This study aimed to investigate whether repetitive optimizing left ventricular pacing configurations (LVPCs) with quadripolar leads (QUAD) can improve response to cardiac resynchronization therapy (CRT). METHODS Fifty-two eligible patients were enrolled and 1:1 randomized to either the quadripolar LV leads (QUAD) group or the conventional bipolar leads (CONV) group. In the QUAD group, optimization of LVPC was performed for all patients before discharge and for nonresponders at 3 months follow-up. Clinical evaluations and transthoracic echocardiograms were performed before, 3, and 6 months after CRT implantation. RESULTS At 3 months follow-up, 16 of 25 (64%) patients in the CONV group (1 patient was lost to follow-up) and 18 of 26 (69%) patients in the QUAD group were classified as responders. After optimizing the LVPCs in 3-month nonresponders in the QUAD group, 21 of 26 (80.8%) patients in the QUAD group were classified as responders at 6 months as compared with 17 of 25 (68%) patients in the CONV group. Left ventricular end-systolic volume (LVESV) reduction, left ventricular ejection fraction (LVEF) increase, and New York Heart Association (NYHA) functional class reduction at 6 months were significantly greater in the QUAD group than in the CONV group (LVESV: -26.9 ± 13.8 vs -17.2 ± 13.3%; P = .013; LVEF: +12.7 ± 8.0 vs +7.8 ± 6.3 percentage points; P = .017; NYHA: -1.27 ± 0.67 vs -0.72 ± 0.54 functional classes; P = .002). CONCLUSIONS Compared with conventional bipolar leads, CRT using quadripolar leads with repetitive optimized LVPCs resulted in an additional increase in LVEF and reduction in LVESV and NYHA functional class at 6-month follow-up.
منابع مشابه
Quadripolar Leads in Cardiac€Resynchronization Therapy
Ma Despite the benefit of cardiac resynchronization therapy (CRT) in patients with heart failure and conduction delay, a considerable number of patients do not respond substantially. Left ventricular lead position is an important factor in response, restricted by the patient’s specific anatomy and local pathophysiology. Quadripolar leads could enhance response to CRT, offering 4 pacing location...
متن کاملLeft ventricular pacing vector selection by novel echo-particle imaging velocimetry analysis for optimization of quadripolar cardiac resynchronization device: a case report
BACKGROUND The availability of pacing configurations offered by quadripolar left ventricular leads could improve patients' response to cardiac resynchronization therapy; however, the selection of an optimal setting remains a challenge. Echo-particle imaging velocimetry has shown that regional anomalies of synchrony/synergy of the left ventricle are related to the alteration, reduction, or suppr...
متن کاملAcute haemodynamic comparison of multisite and biventricular pacing with a quadripolar left ventricular lead.
AIMS Pacing from multiple sites in the left ventricle (LV) may bring about further resynchronization of the diseased heart compared with biventricular (BiV) pacing. We compared acute haemodynamic response (LV dP/dtmax) of multisite and BiV pacing using a quadripolar LV lead. METHODS AND RESULTS In 21 patients receiving cardiac resynchronization therapy, a quadripolar LV lead and conventional ...
متن کاملUse of a quadripolar left ventricular lead to achieve successful implantation in patients with previous failed attempts at cardiac resynchronization therapy
AIMS Problems with implanting a left ventricular (LV) lead during cardiac resynchronization therapy (CRT) procedures are not uncommon and may occur for a variety of reasons including phrenic nerve stimulation (PNS) and high capture thresholds. We aimed to perform successful CRT in patients with previous LV lead problems using the multiple pacing configurations available with the St Jude Quartet...
متن کاملCardiac Resynchronization Therapy Delivered Via a Multipolar Left Ventricular Lead is Associated with Reduced Mortality and Elimination of Phrenic Nerve Stimulation: Long‐Term Follow‐Up from a Multicenter Registry
INTRODUCTION Cardiac resynchronization therapy (CRT) using quadripolar left ventricular (LV) leads provides more pacing vectors compared to bipolar leads. This may avoid phrenic nerve stimulation (PNS) and allow optimal lead placement to maximize biventricular pacing. However, a long-term improvement in patient outcome has yet to be demonstrated. METHODS A total of 721 consecutive patients wi...