Comparative study of retinal nerve fiber layer measurement by StratusOCT and GDx VCC, I: correlation analysis in glaucoma.

نویسندگان

  • Christopher Kai-shun Leung
  • Wai-man Chan
  • Kelvin Kam-Long Chong
  • Wing-ho Yung
  • Kai-tat Tang
  • Jackson Woo
  • Woon-ming Chan
  • Kwok-kay Tse
چکیده

PURPOSE To evaluate the average and regional correlations of retinal nerve fiber layer (RNFL) thickness measured by StratusOCT (optical coherence tomography; Carl Zeiss Meditec, Inc., Dublin, CA) and GDx VCC (Laser Diagnostic Technologies, Inc., San Diego, CA). METHODS Eighty-nine subjects-27 normal, 21 with suspected glaucoma, and 41 with glaucoma-were included in this cross-sectional study. The total average and the mean 12-clock-hour RNFL thickness were measured with the StratusOCT and GDx VCC. The discriminating powers of the two techniques for detection of suspected glaucoma and glaucoma were compared by the area under the receiver operating characteristic curves (AUC). Correspondence between StratusOCT and GDx VCC RNFL measurements in each clock hour was examined with linear regression analysis. RESULTS The average RNFL thickness in the normal group was measured at 101.38 +/- 7.73 and 55.26 +/- 4.32 mum by StratusOCT and GDx VCC, respectively. Both nerve fiber analyzers demonstrated a double-hump pattern in the RNFL profiles with maximum RNFL thickness located at the inferotemporal and superotemporal clock hours by the StratusOCT and the superior and inferior clock hours by the GDx VCC. Significant differences were found in the total average and the individual clock-hour RNFL thickness between StratusOCT and GDx VCC RNFL measurements in both the normal and the suspected glaucoma/glaucoma groups. The GDx VCC superior RNFL measurement demonstrated the largest AUC (0.909) for detection of suspected glaucoma and glaucoma, whereas the largest AUC (0.901) in StratusOCT was found over the inferotemporal clock hour. The total average RNFL thickness measured with StratusOCT and GDx VCC correlated highly with each other (r = 0.852). When the respective clock-hour RNFL measurements were compared, the correlation coefficient varied with the position around the optic nerve head, with the highest correlation found over the superior and inferior clock hours (11, 12, 1, 6, and 7 o'clock; all with r > 0.700) and the lowest located at the temporal clock hour (9 o'clock; r = 0.277). CONCLUSIONS Despite the substantial differences in the values of RNFL thickness, significant correlations were observed between StratusOCT and GDx VCC RNFL measurements. The variations of the correlation coefficient around the optic nerve head suggested that GDx VCC RNFL measurement does not have a fixed relationship with that of StratusOCT and the use of site-specific RNFL birefringences may improve the estimation of RNFL thickness by the GDx VCC. Nevertheless, the GDx VCC was found to be as effective as the StratusOCT in detecting the loss of RNFL in glaucoma.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparative study of retinal nerve fiber layer measurement by StratusOCT and GDx VCC, II: structure/function regression analysis in glaucoma.

PURPOSE To evaluate the structure/function relationship between visual field sensitivity and retinal nerve fiber layer (RNFL) thickness measured by StratusOCT (Carl Zeiss Meditec, Inc., Dublin, CA) and GDx VCC (Laser Diagnostic Technologies, Inc., San Diego, CA). METHODS Eighty-nine subjects (27 who had healthy eyes, 21 who were glaucoma suspect, 41 who had glaucoma) were enrolled in this cro...

متن کامل

Impact of diabetic retinopathy on quantitative retinal nerve fiber layer measurement and glaucoma screening.

PURPOSE To investigate the impact of diabetic retinopathy on quantitative retinal nerve fiber layer (RNFL) assessment and diagnostic power for glaucoma by scanning laser polarimetry (GDx-VCC) and optical coherence tomography (StratusOCT). METHODS The individual RNFL parameters of GDx and OCT were obtained for 170 eyes (one eye from each of 170 subjects [45 healthy, 47 glaucoma, 40 diabetes, a...

متن کامل

Optic disk and nerve fiber layer imaging to detect glaucoma.

PURPOSE To compare optic disk and retinal nerve fiber layer (RNFL) imaging methods to discriminate eyes with early glaucoma from normal eyes. DESIGN Retrospective, cross-sectional study. METHODS In a tertiary care academic glaucoma center, 92 eyes of 92 subjects (46 with early perimetric open-angle glaucoma and 46 controls) were studied. Diagnostic performance of optical coherence tomograph...

متن کامل

Optic disc imaging in perimetrically normal eyes of glaucoma patients with unilateral field loss.

PURPOSE To compare the ability of optic disc and retinal nerve fiber layer (RNFL) imaging to discriminate perimetrically unaffected eyes of glaucoma patients from normal eyes. METHODS Forty-six primary open-angle glaucoma patients with glaucomatous visual field loss with achromatic perimetry in one eye and a normal visual field in the fellow eye and 46 normal controls were selected. Receiver ...

متن کامل

Relationship Between Visual Field Sensitivity and Retinal Nerve Fiber Layer Thickness Measured by Scanning Laser Polarimetry and Optical Coherence Tomography in Normal, Ocular Hypertensive and Glaucomatous Eyes

PURPOSE: To evaluate the correlation between automated achromatic perimetry (AAP) and the output of two retinal nerve fiber layer (RNFL) analysers: scanning laser polarimetry (GDx-VCC) and optical coherence tomography (OCT). METHODS: Quantitative RNFL measurements with GDx-VCC and Stratus-OCT were obtained in one eye from 52 healthy subjects, 38 ocular hypertensive (OHT) patients and 94 glaucom...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Investigative ophthalmology & visual science

دوره 46 9  شماره 

صفحات  -