Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma
نویسندگان
چکیده
The objective of this study was to examine the analytical performance of 14 comprehensive metabolic panel analytes on the Abaxis Piccolo Xpress® Point of Care analyzer in serum, plasma, and whole blood. A method comparison was performed on all three specimen types intended for use on the Piccolo Xpress®: serum, heparinized plasma, and whole blood. This data is also presented in Murata et al. (2015) [1]. This article includes the actual Bland-Altman bias plots of the difference in results obtained for analytes in the comprehensive metabolic panel from the Abaxis Piccolo Xpress and the comparison instrument, the Ortho Vitros.
منابع مشابه
Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital
OBJECTIVES Point of care testing (POCT) contributes to diagnosis and monitoring with fast testing time and easily performed assays. We evaluated the Abaxis Piccolo Xpress point of care chemistry analyzer using the Liver Panel Plus discs for our pediatric patient population at Texas Children's Hospital. DESIGN AND METHODS Analytical performance was evaluated for precision and linearity using q...
متن کامل[Performance evaluation of the Piccolo xpress Point-of-care Chemistry Analyzer].
BACKGROUND Point-of-care (POC) tests are used increasingly due to fast results and simple test procedures, which enables rapid diagnosis and therapeutic monitoring. We evaluated the performance of the Piccolo xpress Chemistry Analyzer (Abaxis, USA) a POC chemistry analyzer. METHODS Fourteen analytes, Na(+), K(+), Cl(-), Ca(2+), total carbon dioxide, AST, ALT, total bilirubin, alkaline phospha...
متن کاملPerformance of an Optimized Paper-Based Test for Rapid Visual Measurement of Alanine Aminotransferase (ALT) in Fingerstick and Venipuncture Samples
BACKGROUND A paper-based, multiplexed, microfluidic assay has been developed to visually measure alanine aminotransferase (ALT) in a fingerstick sample, generating rapid, semi-quantitative results. Prior studies indicated a need for improved accuracy; the device was subsequently optimized using an FDA-approved automated platform (Abaxis Piccolo Xpress) as a comparator. Here, we evaluated the pe...
متن کاملPiccolo® Comprehensive Metabolic Reagent Disc
chemistry analyzer, is intended to be used for the in vitro quantitative determination of alanine aminotransferase (ALT), albumin, alkaline phosphatase (ALP), aspartate aminotransferase (AST), calcium, chloride, creatinine, glucose, potassium, sodium, total bilirubin, total carbon dioxide, total protein, and blood urea nitrogen (BUN) in heparinized whole blood, heparinized plasma, or serum in a...
متن کاملEvaluation of the Nova StatSensor® XpressTM Creatinine Point-Of-Care Handheld Analyzer
Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-limited settings, where more instruments that are sophisticated are not available. The StatSensor X...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 16 شماره
صفحات -