Impact of interfraction seroma collection on breast brachytherapy dosimetry – a mathematical model
نویسندگان
چکیده
PURPOSE Balloon brachytherapy is a widely accepted modality for delivery of accelerated partial breast irradiation (APBI). Our hypothesis was that inter-fraction seroma collection around the balloon surface would have an adverse effect on dosimetry of the target. MATERIAL AND METHODS This is a dosimetric re-planning study using two volumetric models (30 cc and 45 cc) in a Contura(®) multi-lumen balloon (MLB) catheter. In a previously treated patient, two customized baseline plans were generated using multiple channels of the Contura(®) catheter prescribed to the Planning Target Volume Evaluation (PTV_Eval). Symmetric expansions of 1.0 mm (0-9 mm) increments around the balloon surface were performed to simulate a "Virtual Seroma" (VS) accumulation for both balloon volumes and plans were obtained for each expansion using Eclipse Brachyvision™. An analysis of these plans was then performed to evaluate the effect of seroma accumulation on dosimetric parameters of V100 and V90. RESULTS 20 plans were generated and analyzed (10 plans for each balloon volume), representing VS of 6.0-66.0 cc. There was a commensurate decrease in the dose delivered to the PTV_Eval V100 and V90 (as defined by the original treatment plan) with increasing VS accumulation leading to a sub-optimal coverage of the PTV_Eval. For 30 cc MLB catheter, V100 decreased by 1.4% and V90 decreased by 0.9% for every 1 cc of VS. For 45cc MLB catheter, V100 decreased by 1.3% and V90 decreased by 1.15% for every 1.0 cc accumulation of VS. CONCLUSIONS Balloon catheter-tissue adherence ensures daily dose delivery to the planned PTV_Eval. Accumulation of seroma, hematoma or air between HDR fractions can significantly impact PTV_Eval dosimetry. Vacuum-port aspiration prior to delivery of each fraction, if available, should be considered to minimize the risk of geographic under dosing.
منابع مشابه
A study on dosimetry accuracy of Strut-Adjust Volume Implant (SAVI) brachytherapy
Introduction: Accelerated Partial Breast Irradiation (APBI) is an effective treatment for breast carcinomas because of its limited number of fractions. Strut-adjusted volume implant (SAVI) is a HDR brachytherapy applicator to deliver the prescription dose with good PTV coverage and acceptable dose to skin, chest wall and organ at risks. Acceptable clinical outcomes depend on a...
متن کاملMonte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost
Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phan...
متن کاملThe Effect of Breast Phantom Material on the Dose Distribution in AccuBoost Brachytherapy
Introduction: Long-term teletherapy program is not suitable for old and working patients and those living in areas where little access to primary health care is available. Accelerated partial breast irradiation (APBI) using high dose rate (HDR) brachytherapy is an appropriate alternative for these patients due to its limited number of fractions. The AccuBoost is a system for delivering APBI. Th...
متن کاملA Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy
Background: Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy.Objective: To investigate the percentage of the dose received by critical organ (skin), thermoluminescence detec...
متن کاملTreatment plan dosimetric impact after catheter stretch during multi-catheter interstitial breast brachytherapy
Purpose Nylon 6/6 interstitial brachytherapy catheters may stretch when exposed to moisture, mechanical tension, and body temperature. The purpose of this study is to evaluate the magnitude of catheter stretch during a course of multi-catheter interstitial breast brachytherapy for nylon 6/6 treatment catheters, and to assess the impact this has on treatment plan dosimetry. Material and method...