A hierarchical approach for online temporal lobe seizure detection in long-term intracranial EEG recordings.
نویسندگان
چکیده
OBJECTIVE Around 1% of the world's population is affected by epilepsy, and nearly 25% of patients cannot be treated effectively by available therapies. The presence of closed-loop seizure-triggered stimulation provides a promising solution for these patients. Realization of fast, accurate, and energy-efficient seizure detection is the key to such implants. In this study, we propose a two-stage on-line seizure detection algorithm with low-energy consumption for temporal lobe epilepsy (TLE). APPROACH Multi-channel signals are processed through independent component analysis and the most representative independent component (IC) is automatically selected to eliminate artifacts. Seizure-like intracranial electroencephalogram (iEEG) segments are fast detected in the first stage of the proposed method and these seizures are confirmed in the second stage. The conditional activation of the second-stage signal processing reduces the computational effort, and hence energy, since most of the non-seizure events are filtered out in the first stage. MAIN RESULTS Long-term iEEG recordings of 11 patients who suffered from TLE were analyzed via leave-one-out cross validation. The proposed method has a detection accuracy of 95.24%, a false alarm rate of 0.09/h, and an average detection delay time of 9.2 s. For the six patients with mesial TLE, a detection accuracy of 100.0%, a false alarm rate of 0.06/h, and an average detection delay time of 4.8 s can be achieved. The hierarchical approach provides a 90% energy reduction, yielding effective and energy-efficient implementation for real-time epileptic seizure detection. SIGNIFICANCE An on-line seizure detection method that can be applied to monitor continuous iEEG signals of patients who suffered from TLE was developed. An IC selection strategy to automatically determine the most seizure-related IC for seizure detection was also proposed. The system has advantages of (1) high detection accuracy, (2) low false alarm, (3) short detection latency, and (4) energy-efficient design for hardware implementation.
منابع مشابه
Preictal state identification by synchronization changes in long-term intracranial EEG recordings.
OBJECTIVE There is accumulated evidence that mesial temporal lobe seizures are preceded by a preictal transition that evolves over minutes to hours. In the present study, we investigated these possible preictal changes in long-term intracranial recordings of five patients by a measure of phase synchronization. In order to clearly distinguish preictal changes from all the other interictal states...
متن کاملPerformance of a seizure warning algorithm based on the dynamics of intracranial EEG.
During the past decade, several studies have demonstrated experimental evidence that temporal lobe seizures are preceded by changes in dynamical properties (both spatial and temporal) of electroencephalograph (EEG) signals. In this study, we evaluate a method, based on chaos theory and global optimization techniques, for detecting pre-seizure states by monitoring the spatio-temporal changes in ...
متن کاملElectroencephalography in Mesial Temporal Lobe Epilepsy: A Review
Electroencephalography (EEG) has an important role in the diagnosis and classification of epilepsy. It can provide information for predicting the response to antiseizure drugs and to identify the surgically remediable epilepsies. In temporal lobe epilepsy (TLE) seizures could originate in the medial or lateral neocortical temporal region, and many of these patients are refractory to medical tre...
متن کاملApplication of a multivariate seizure detection and prediction method to non-invasive and intracranial long-term EEG recordings.
OBJECTIVE Retrospective evaluation and comparison of performances of a multivariate method for seizure detection and prediction on simultaneous long-term EEG recordings from scalp and intracranial electrodes. METHODS Two multivariate techniques based on simulated leaky integrate-and-fire neurons were investigated in order to detect and predict seizures. Both methods were applied and assessed ...
متن کاملDC-EEG recording: a paradigm shift in seizure localization?
Interictal or ictal discharges are the basis of EEG interpretation when seizure localization is the goal, but a new report suggests an alternative. It has been known for over 40 years that, when DC recordings are made, seizures may be associated with localized, very slow, sustained voltage changes, which do not oscillate (ictal DC shift).1 These voltage changes occur more or less simultaneously...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of neural engineering
دوره 10 4 شماره
صفحات -
تاریخ انتشار 2013