Re-usage of external pacing electrodes results in skin burns.

نویسندگان

  • Alexander Loch
  • Choon Chin Ang
  • Wan Azman Wan Ahmad
چکیده

A 29 year old woman with a background of poorly controlled iabetes was admitted to our hospital in cardiogenic shock due o acute inferior myocardial infarction and complete heart block. ecycled external electrodes (“stat.padz®”) were placed to the right f the upper sternal border below the clavicle and to the left of he nipple with the centre of the electrode in the mid-axillary line. xternal pacing was commenced with an output of 90 mA at a rate f 70 stimulations per minute. Mechanical capture was confirmed y echocardiography. Primary percutaneous coronary intervention o the right coronary artery was performed with insertion of a ransvenous temporary pacemaker. The patient developed intrarocedurally ventricular fibrillation for which a shock of 200 J was elivered through the external electrodes. The total duration of xternal pacing was about 90 min. Upon removal of the external lectrodes, multiple 2nd degree circular skin burns were noticed elow the apical electrode with burnt and necrotic skin flakes dherent to the electrode (Panel A). No burn was observed in the rea of the right sided electrode (see Fig. 1). Many of our patients do not have medical insurance nor are they ble to pay their hospital bills. The price for a set of external pacing lectrodes is about 50 USD in our setting. Every patient requiring ife-saving treatment will be treated regardless of insurance status. his policy however results frequently in the hospital not recoverng cost for consumable products like pacing electrodes. In an effort o reduce cost, external electrodes are re-used – despite the manfacturer’s advice. The gel of re-used electrodes is often dried out ith pockets of gel missing in addition to general contamination nd cracks (Panel B). We assume that the relatively severe burns

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Burns to be alive: a complication of transcutaneous cardiac stimulation

Figure 1 Two little third-degree skin burns under the breast following use of an external cardiac pacing device. External cardiac pacing devices (ECPDs) are commonly employed in emergency situations. The main indication for their use is untolerated bradycardia, especially in cases where other medical treatments have no effect [1]. Although the efficacy and safety of ECPDs are well documented, a...

متن کامل

Evaluation of chitosan gel on burn wound healing and keratinocytes function

Skin burns are common skin injuries that can result from exposure to various sources of heat by which, depending on the cause, different degrees of burns are generated. In restoration of superficial burns by re-epithelialization, the skin is rebuilt, but the skin color is different than the healthy tissue. Keratinocytes represent the major cells of the epidermis. These cells are sourced from ec...

متن کامل

External cardiac pacing using low impedance electrodes suitable for defibrillation: a comparative blinded study.

OBJECTIVES The objective of this study was to determine whether the threshold for successful cardiac pacing is affected by electrode impedance and whether this procedure can be successfully carried out through low impedance electrodes that are also suitable for defibrillation. BACKGROUND Reintroduced in the early 1980s, external cardiac pacing utilizes large externally placed electrodes with ...

متن کامل

بررسی تأثیر امواج فرابنفش ناشی از جوشکاری با الکترودهای صنعتی در القای سرطان پوست در موشهای سوری BaIb/c

All of the ultraviolet wave radiations, including UVA can alter the immune system of the skin. Scale cells cancer of the skin and malignant melanoma are caused by UVA radiation, but researchers insist more on the carcinogenetic effect of UVB spectrum. UVB spectrum is produced from the sun, sodium vapor lamp and electric and welding radiation of metal electrodes. In this study, the effect of UV ...

متن کامل

Types of pacemakers and their complications.

1. Implantable pulse generators with endocardial or myocardial electrodes for long-term or permanent use. 2. External, miniaturized, transistorized, patientportable, battery-powered, pulse generators with exteriorized electrodes for temporary transvenous endocardial or transthoracic myocardial pacing. 3. Console batteryor AC-powered cardiovertors, defibrillators, or monitors with high-current e...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Resuscitation

دوره 84 9  شماره 

صفحات  -