Multifocal visual evoked potential is dependent on electrode position.
نویسندگان
چکیده
PURPOSE To investigate whether the multifocal visual evoked potential (mVEP) is dependent on the electrode position, and to confirm the reproducibility of the mVEP. METHODS The mVEPs were recorded using the Veris III system with two different bipolar electrode settings. In Position 1, electrodes were placed at equal distances in vertical alignment 2 cm above and below the inion. In Position 2, the electrodes were placed in horizontal alignment at equal distances 2 cm to the left and right of the inion. Dartboard pattern stimulation was conducted. The mVEPs were repeatedly recorded from 4 volunteers, and mainly the second-order kernel response components were analyzed. RESULTS Although the reproducibility of mVEP was good in both Position 1 and Position 2, each waveform of the mVEP was drastically different between the two positions. This difference in the waveforms was clearly shown in the center and at the horizontal meridian. We also investigated the first-order kernel response components of the mVEPs. Several traces of the first-order kernel response components did not reveal flat traces. This point is also worthy of consideration. CONCLUSION Responses from mVEPs are clearly dependent on the electrode position.
منابع مشابه
Multifocal topographic visual evoked potential: improving objective detection of local visual field defects.
PURPOSE To investigate the relationships between the pattern stimulation of different parts of the visual field (up to 25 degrees of eccentricity), the electrode position, and the cortical response to improve objective detection of local visual field defects. METHODS The human visual evoked potential (VEP) was assessed using multifocal pseudorandomly alternated pattern stimuli that were corti...
متن کاملOptimizing electrode positions and analysis strategies for multifocal VEP recordings by ROC analysis
The multifocal visual evoked potential (mfVEP) is an important tool to test visual pathway function. The aim of this study was to optimize electrode positions in mfVEP recordings. For analysis we applied a receiver operating characteristic (ROC), a method that inherently corrects for multiple testing. We found that a combination of two perpendicular derivations-both straddling the inion-was the...
متن کاملComparing multifocal pupillographic objective perimetry (mfPOP) and multifocal visual evoked potentials (mfVEP) in retinal diseases
Multifocal pupillographic objective perimetry (mfPOP) shows regions of slight hypersensitivity away from retinal regions damaged by diabetes or age-related macular degeneration (AMD). This study examines if such results also appear in multifocal visual evoked potentials (mfVEPs) recorded on the same day in the same patients. The pupil control system receives input from the extra-striate cortex,...
متن کاملComparison of visual evoked potential changes in patients with alzheimer, vascular dementia and minimal cognitive Impairment with healthy people: a case-control study
Background: Alzheimer dementia as the most common cause of dementia is a chronic, progressive, irreversible and incurable disease. The second most common cause of dementia after Alzheimer is vascular dementia. One of the systems involved in dementia is the visuospatial system and visual evoked potential (VEP) can be one of the diagnostic methods for this disease. Therefore, the present study ai...
متن کاملMultifocal visual evoked potential in nonorganic visual field loss.
OBJECTIVE To evaluate the use of multifocal visual evoked potentials in the diagnosis of nonorganic visual field loss. METHODS Five patients with unexplained visual field loss underwent full neuro-ophthalmic examination, Humphrey visual field testing, and multifocal visual evoked potential testing using the Accumap objective perimeter (ObjectiVision Pty Ltd, Sydney, Australia). RESULTS In a...