Reconstruction of type II pelvic resection with a new peri-acetabular reconstruction endoprosthesis.
نویسندگان
چکیده
We report our early experience with a new peri-acetabular reconstruction endoprosthesis used for pelvic reconstruction after tumour resection. The outcome of 21 patients who underwent limb salvage following type II pelvic resection and reconstruction using the peri-acetabular reconstruction prosthesis between 2000 and 2006 was retrospectively reviewed. This prosthesis was designed to use the remaining part of the ilium to support a horizontally placed acetabular component secured with internal fixation and bone cement. Into this device a constrained acetabular liner is positioned which is articulates with a conventional femoral component to which a modular extension and modular head are attached. The mean follow-up was 20.5 months (1 to 77). The most common complications were deep infection, superficial wound infections, and dislocation. The mean musculoskeletal tumor society functional outcome score for the survivors was 20.1(11 to 27). We recommend the use of the peri-acetabular reconstruction prosthesis for reconstruction of large defects after type II pelvic resection, as this design has a greater inherent stability over other available prostheses.
منابع مشابه
Clinical outcome of pedestal cup endoprosthetic reconstruction after resection of a peri-acetabular tumour.
Peri-acetabular tumour resections and their subsequent reconstruction are among the most challenging procedures in orthopaedic oncology. Despite the fact that a number of different pelvic endoprostheses have been introduced, rates of complication remain high and long-term results are mostly lacking. In this retrospective study, we aimed to evaluate the outcome of reconstructing a peri-acetabula...
متن کاملPeri-acetabular resection and endoprosthetic reconstruction for tumours of the acetabulum.
We treated 98 patients with peri-acetabular tumours by resection and reconstruction with a custom-made pelvic endoprosthesis. The overall survival of the patients was 67% at five years, 54% at ten years and 51% at 30 years. One or more complications occurred in 58.1% of patients (54), of which infection was the most common, affecting 30% (28 patients). The rate of local recurrence was 31% (29 p...
متن کاملReconstruction with Modular Hemipelvic Endoprosthesis after Pelvic Tumor Resection: A Report of 50 Consecutive Cases
PURPOSE To evaluate the effectiveness of reconstruction with a modular hemipelvic endoprosthesis after pelvic tumor resection. METHODS We retrospectively studied 50 consecutive patients diagnosed with pelvic tumor from 2003 to 2013. All patients received limb-salvage surgery and reconstruction with modular hemipelvic endoprosthesis. RESULTS Patients were followed for an average of 54 months...
متن کامل10730- Massive pelvic allograft reconstruction of the acetabular area: management of hip mechanical failure
Introduction: Reconstruction of the hip after the resection of a pelvic tumor is a challenging procedure in orthopaedic oncology. The use of massive allograft reconstruction may restore the pelvic architecture and provide a suitable bone stock for acetabular prosthetic implant. The major complication is deep infection whatever the method of reconstruction applied; however, high mechanical compl...
متن کاملComplex reconstruction in the management of extremity sarcomas.
The concept of limb-sparing surgery for bony sarcomas has evolved over the past 25 years. Today, more than 90% of patients treated by surgeons with expertise in musculoskeletal oncology undergo successful limb-sparing procedures. Many large centers have abandoned osteochondral allografts and resection arthrodesis for the reconstruction of segmental bone and joint defects in favor of metallic en...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 90 3 شماره
صفحات -
تاریخ انتشار 2008