An iterative approach to the beam hardening correction in cone beam CT.
نویسندگان
چکیده
In computed tomography (CT), the beam hardening effect has been known to be one of the major sources of deterministic error that leads to inaccuracy and artifact in the reconstructed images. Because of the polychromatic nature of the x-ray source used in CT and the energy-dependent attenuation of most materials, Beer's law no longer holds. As a result, errors are present in the acquired line integrals or measurements of the attenuation coefficients of the scanned object. In the past, many studies have been conducted to combat image artifacts induced by beam hardening. In this paper, we present an iterative beam hardening correction approach for cone beam CT. An algorithm that utilizes a tilted parallel beam geometry is developed and subsequently employed to estimate the projection error and obtain an error estimation image, which is then subtracted from the initial reconstruction. A theoretical analysis is performed to investigate the accuracy of our methods. Phantom and animal experiments are conducted to demonstrate the effectiveness of our approach.
منابع مشابه
Artifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality
Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...
متن کاملMulti-Material Beam Hardening Correction(MMBHC) in Computed Tomography
In computed tomography (CT), the nonlinear attenuation characteristics of polychromatic X-rays cause beam hardening artifacts in the reconstructed images. State-of-theart methods to correct the beam hardening effect are mostly single material precorrections (e.g. water-precorrection), which are far less efficient when more than one material is present in the field of measurement. The use of tho...
متن کاملSpectrum Binning Approach for Multi-Material Beam Hardening Correction (MMBHC) in CT
In CT, the nonlinear attenuation characteristics of polychromatic X-rays cause beam hardening artifacts in the reconstructed images. When datasets contain multiple materials in the field of scan, the performance of state-of-the-art beam hardening correction approaches is limited by high computational load and reduced correction efficiency. An image based multi-material beam hardening correction...
متن کاملBeam Hardening Correction in X-Ray Computed Tomography: A Comparison of Two Iterative Model-Based Reconstruction Methods
While the expenses for computational power decrease, iterative reconstruction methods for x-ray computed tomography (CT) that allow the use of improved model assumptions, become more attractive. In applications for non-destructive testing one of the most common degradations of image quality with standard x-ray CT reconstruction methods is beam hardening. Techniques for beam hardening correction...
متن کاملBeam Hardening Correction with an Iterative Scheme Using an Exact Backward Projector and a Polychromatic Forward Projector
In computed tomography (CT), reconstructions from conebeam (CB) data acquired with a polychromatic X-ray device show so called beam hardening artifacts. Beam hardening artifacts are highly undesirable for a medical diagnosis, because details in the reconstructed image are severely disturbed or completely lost. In this work, we demonstrate the significant reduction of beam hardening artifacts by...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Medical physics
دوره 27 1 شماره
صفحات -
تاریخ انتشار 2000