Accuracy verification of infrared marker-based dynamic tumor-tracking irradiation using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000).

نویسندگان

  • Nobutaka Mukumoto
  • Mitsuhiro Nakamura
  • Akira Sawada
  • Yasunobu Suzuki
  • Kunio Takahashi
  • Yuki Miyabe
  • Shuji Kaneko
  • Takashi Mizowaki
  • Masaki Kokubo
  • Masahiro Hiraoka
چکیده

PURPOSE To verify the accuracy of an infrared (IR) marker-based dynamic tumor-tracking irradiation system (IR tracking) using the gimbaled x-ray head of the Vero4DRT (MHI-TM2000). METHODS The gimbaled 6-MV C-band x-ray head of the Vero4DRT can swing along the pan-and-tilt direction to track a moving target. During beam delivery, the Vero4DRT predicts the future three-dimensional (3D) target position in real time using a correlation model [four-dimensional (4D) model] between the target and IR marker motion, and then continuously transfers the corresponding tracking orientation to the gimbaled x-ray head. The 4D-modeling error (E4DM) and the positional tracking error (EP) were defined as the difference between the predicted and measured positions of the target in 4D modeling and as the difference between the tracked and measured positions of the target during irradiation, respectively. For the clinical application of IR tracking, we assessed the relationship between E4DM and EP for three 1D sinusoidal (peak-to-peak amplitude [A]: 20-40 mm, breathing period [T]: 2-4 s), five 1D phase-shifted sinusoidal (A: 20 mm, T: 4 s, phase shift [τ]: 0.2-2 s), and six 3D patient respiratory patterns. RESULTS The difference between the 95th percentile of the absolute EP (EP (95)) and the mean (μ) + two standard deviations (SD) of absolute E4DM (E4DM (μ+2SD)) was within ± 1 mm for all motion patterns. As the absolute correlation between the target and IR marker motions decreased from 1.0 to 0.1 for the 1D phase-shifted sinusoidal patterns, the E4DM (μ+2SD) and EP (95) increased linearly, from 0.4 to 3.0 mm (R = -0.98) and from 0.5 to 2.2 mm (R = -0.95), respectively. There was a strong positive correlation between E4DM (μ+2SD) and EP (95) in each direction [(lateral, craniocaudal, anteroposterior) = (0.99, 0.98, 1.00)], even for the 3D respiratory patterns; thus, EP (95) was readily estimated from E4DM (μ+2SD). CONCLUSIONS Positional tracking errors correlated strongly with 4D-modeling errors in IR tracking. Thus, the accuracy of the 4D model must be verified before treatment, and margins are required to compensate for the 4D-modeling error.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Positional accuracy of novel x-ray-image-based dynamic tumor-tracking irradiation using a gimbaled MV x-ray head of a Vero4DRT (MHI-TM2000).

PURPOSE To verify the positional accuracy of a novel x-ray-image-based dynamic tumor-tracking (DTT) irradiation technique using the gimbaled MV x-ray head of a Vero4DRT (MHI-TM2000). METHODS Verification of the x-ray-image-based DTT was performed using three components: a three-dimensional moving phantom with a steel ball target, a laser displacement gauge, and an orthogonal kV x-ray imaging ...

متن کامل

Effect of audio instruction on tracking errors using a four‐dimensional image‐guided radiotherapy system

The Vero4DRT (MHI-TM2000) is capable of performing X-ray image-based tracking (X-ray Tracking) that directly tracks the target or fiducial markers under continuous kV X-ray imaging. Previously, we have shown that irregular respiratory patterns increased X-ray Tracking errors. Thus, we assumed that audio instruction, which generally improves the periodicity of respiration, should reduce tracking...

متن کامل

Feasibility evaluation of a new irradiation technique: three-dimensional unicursal irradiation with the Vero4DRT (MHI-TM2000)

The Vero4DRT (MHI-TM2000) is a newly designed unique image-guided radiotherapy system consisting of an O-ring gantry. This system can realize a new irradiation technique in which both the gantry head and O-ring continuously and simultaneously rotate around the inner circumference of the O-ring and the vertical axis of the O-ring, respectively, during irradiation. This technique creates three-di...

متن کامل

Long‐term stability assessment of a 4D tumor tracking system integrated into a gimbaled linear accelerator

We assessed long-term stability of tracking accuracy using the Vero4DRT system. This metric was observed between September 2012 and March 2015. A programmable respiratory motion phantom, designed to move phantoms synchronously with respiratory surrogates, was used. The infrared (IR) markers moved in the anterior-posterior (AP) direction as respiratory surrogates, while a cube phantom with a ste...

متن کامل

Baseline correction of a correlation model for improving the prediction accuracy of infrared marker‐based dynamic tumor tracking

We previously found that the baseline drift of external and internal respiratory motion reduced the prediction accuracy of infrared (IR) marker-based dynamic tumor tracking irradiation (IR Tracking) using the Vero4DRT system. Here, we proposed a baseline correction method, applied immediately before beam delivery, to improve the prediction accuracy of IR Tracking. To perform IR Tracking, a four...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 40 4  شماره 

صفحات  -

تاریخ انتشار 2013