Report of the IFCC Working Group for Standardization of Thyroid Function Tests; part 3: total thyroxine and total triiodothyronine.

نویسندگان

  • Linda M Thienpont
  • Katleen Van Uytfanghe
  • Graham Beastall
  • James D Faix
  • Tamio Ieiri
  • W Greg Miller
  • Jerald C Nelson
  • Catherine Ronin
  • H Alec Ross
  • Jos H Thijssen
  • Brigitte Toussaint
چکیده

BACKGROUND Because total thyroid hormone testing is performed on many automated clinical chemistry instruments, the IFCC Scientific Division commissioned the Working Group for Standardization of Thyroid Function Tests to include total thyroxine (TT4) and total triiodothyronine (TT3) in its standardization efforts. METHODS Existing SI-traceable reference measurement procedures (RMPs) were used to assign TT4 and TT3 values to 40 single-donor serum samples for subsequent use in a method comparison study with 11 TT4 and 12 TT3 immunoassays. Data from comparison of each immunoassay with the RMPs provided a basis for mathematical assay recalibration. RESULTS Seven TT4 assays had a mean bias within 10% of the RMP, but 2 deviated by an average of -12% and another 2 by +17%. All TT3 assays showed positive biases, 4 within and 8 outside 10%, up to 32%. Mathematical recalibration effectively eliminated assay-specific biases, but sample-related effects remained, particularly for TT3. Correlation coefficients with the RMPs ranged from 0.82 to 0.97 for TT4 and from 0.32 to 0.92 for TT3. The within-run and total imprecision ranges for TT4 were 1.4% to 9.1% and 3.0% to 9.4%, respectively, and for TT3 2.1% to 7.8% and 2.8% to 12.7%, respectively. Approximately one-half of the assays matched the internal QC targets within approximately 5%; however, we observed within-run drifts/shifts. CONCLUSIONS The study showed that of the assays we examined, only 4 TT4 but the majority of the TT3 assays needed establishment of calibration traceability to the existing RMPs. Most assays performed well, but some would benefit from improved precision, within-run stability, and between-run consistency.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The effects of long term handling stress on thyroid function in male rats

Inappropriate responsiveness of the stress system may cause numerous alterations in endocrine, metabolic and autonomic systems. Many studies have been carried out on the effects of acute stress on thyroid function but the effects of chronic stress have not been fully investigated. The purpose of this study is to evaluate the effects of long term handling stress on thyroid function. Study was pe...

متن کامل

The effects of long term handling stress on thyroid function in male rats

Inappropriate responsiveness of the stress system may cause numerous alterations in endocrine, metabolic and autonomic systems. Many studies have been carried out on the effects of acute stress on thyroid function but the effects of chronic stress have not been fully investigated. The purpose of this study is to evaluate the effects of long term handling stress on thyroid function. Study was pe...

متن کامل

Harmonization and standardization of thyroid function tests.

This issue of Clinical Chemistry contains 3 reports from the IFCC Working Group for Standardization of Thyroid Function Tests. The first report relates to thyroidstimulating hormone (TSH) (1 ), the second addresses free thyroxine (FT4) and free triiodothyronine (FT3) (2 ), and the third discusses total thyroxine (T4) and total triiodothyronine (T3) (3 ). Each of these reports is authored by a g...

متن کامل

Report of the IFCC Working Group for Standardization of Thyroid Function Tests; part 2: free thyroxine and free triiodothyronine.

BACKGROUND Free thyroxine (FT4) and free triiodothyronine (FT3) measurements are useful in the diagnosis and treatment of a variety of thyroid disorders. The IFCC Scientific Division established a Working Group to resolve issues of method performance to meet clinical requirements. METHODS We compared results for measurement of a panel of single donor sera using clinical laboratory procedures ...

متن کامل

Positive interference in Radioimmunoassay determination of serum total T3

Introduction: Determination of thyrotropin, total and free thyroxine and triiodothyronine are widely used as diagnostic methods for thyroid function evaluation. There have been numerous reports of interferences in thyroid hormone immunoassays. Herein, a prominent positive interference is reported for a radioimmunoassay kit of total triiodothyronine. Materials and Methods: In this study, a total...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 56 6  شماره 

صفحات  -

تاریخ انتشار 2010