T2-prepared SSFP improves diagnostic confidence in edema imaging in acute myocardial infarction compared to turbo spin echo.

نویسندگان

  • Peter Kellman
  • Anthony H Aletras
  • Christine Mancini
  • Elliot R McVeigh
  • Andrew E Arai
چکیده

T2-weighted MRI of edema in acute myocardial infarction (MI) provides a means of differentiating acute and chronic MI, and assessing the area at risk of infarction. Conventional T2-weighted imaging of edema uses a turbo spin-echo (TSE) readout with dark-blood preparation. Clinical applications of dark-blood TSE methods can be limited by artifacts such as posterior wall signal loss due to through-plane motion, and bright subendocardial artifacts due to stagnant blood. Single-shot imaging with a T2-prepared SSFP readout provides an alternative to dark-blood TSE and may be conducted during free breathing. We hypothesized that T2-prepared SSFP would be a more reliable method than dark-blood TSE for imaging of edema in patients with MI. In patients with MI (22 acute and nine chronic MI cases), T2-weighted imaging with both methods was performed prior to contrast administration and delayed-enhancement imaging. The T2-weighted images using TSE were nondiagnostic in three of 31 cases, while six additional cases rated as being of diagnostic quality yielded incorrect diagnoses. In all 31 cases the T2-prepared SSFP images were rated as diagnostic quality, correctly differentiated acute or chronic MI, and correctly determined the coronary territory. Free-breathing T2 prepared SSFP provides T2-weighted images of acute MI with fewer artifacts and better diagnostic accuracy than conventional dark-blood TSE.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

SSFP cine images early post-gadolinium improve detection of myocardial oedema in acute myocardial infarction compared to T2-weighted dark blood turbo spin echo images

Background Demonstration of myocardial oedema following MI allows differentiation of acute and chronic MI, and assessment of the area at risk of infarction. T2-weighted dark blood TSE imaging for oedema is technically demanding and prone to artefacts. Our protocol for patients with acute MI include standard oedema imaging sequences and more recently T2 prepared single shot SSFP sequences, follo...

متن کامل

ACUT2E TSE-SSFP: a hybrid method for T2-weighted imaging of edema in the heart.

ACUT(2)E TSE-SSFP is a hybrid between steady state free precession (SSFP) and turbo spin echo (TSE) for bright-blood T2-weighted imaging with signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) similar to dark-blood TSE. TSE-SSFP uses a segmented SSFP readout during diastole with 180 degrees pulses following a 90 degrees preparation. The 180 degrees refocusing pulses make TSE-SSFP sim...

متن کامل

T2-prepared SSFP Improves Diagnostic Confidence in Edema Imaging in Acute Myocardial Infarction Compared with Turbo-SpinEcho

Introduction T2-weighted MR imaging of edema in acute myocardial infarction (MI) provides a means of differentiating acute and chronic MI and for assessing the area-at-risk of infarction. Standard T2-weighted imaging of edema uses a turbo-SpinEcho (TSE) readout with dark-blood preparation. Dark-blood TSE methods are subject to artifacts such as posterior wall signal loss due to cardiac motion a...

متن کامل

Improved T2-Weighted Cardiac Imaging using Retrospective Motion Correction and Optimal Image Combination

Introduction T2-weighted (T2w) cardiac MR imaging has established reputation as a sensitive technique to depict myocardial edema related to acute infarction. Together with delayed enhancement technique, T2w demonstrated valuable clinical potentials to differentiate acute damage from chronic infarction [1]. To minimize the myocardial signal loss caused by through-plane motion, T2w can be perform...

متن کامل

T2 quantification for improved detection of myocardial edema

BACKGROUND T2-Weighted (T2W) magnetic resonance imaging (MRI) pulse sequences have been used to detect edema in patients with acute myocardial infarction and differentiate acute from chronic infarction. T2W sequences have suffered from several problems including (i) signal intensity variability caused by phased array coils, (ii) high signal from slow moving ventricular chamber blood that can mi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 57 5  شماره 

صفحات  -

تاریخ انتشار 2007