Ferromagnetism and MR imaging: safety of carotid vascular clamps.

نویسندگان

  • G P Teitelbaum
  • M C Lin
  • A T Watanabe
  • J F Norfray
  • T I Young
  • W G Bradley
چکیده

Metallic extracranial carotid vascular clamps of the Selverstone, Crutchfield, Poppen-Blaylock, Salibi, Kindt, and tantalum varieties have been placed for treatment of large, giant, or inoperable intracranial aneurysms. To ascertain what adverse effect, if any, MR imaging would have on these clamps, magnetic deflection at 1.5 T was measured for various carotid clamps. Marked magnetic deflection (and torque) was displayed by stainless steel Poppen-Blaylock clamps. Relatively mild magnetic deflection was displayed by the stainless steel Selverstone, Salibi, Crutchfield, and Kindt clamps. Three patients with previously placed carotid clamps (two Selverstone, one Salibi) and one patient with a nonferromagnetic tantalum carotid clip had cranial or cervical MR studies at field strengths ranging from 0.35 to 0.60 T. No patient experienced any discomfort or neurologic sequelae as a result of MR imaging. Although the ferromagnetic clamps created severe "black-hole" artifacts and image distortion within the cervical and facial regions, no significant image degradation was apparent during spin-echo imaging of the brain. The tantalum clip created a far smaller MR artifact than did ferromagnetic clamps and allowed effective spin-echo and gradient-echo imaging in the cervical region. Our findings indicate that most patients with carotid vascular clamps (and nonferromagnetic clips) can probably be imaged safely with MR.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

P1: Dextran Curcumin Promotes Novel Object Recognition Memory in Rats after Ischemic Stroke

Ischemic stroke causes the depletion of energy and induces excitotoxicity and neuroinflammation in the brain that results from thrombotic blockage. Cerebral ischemia leads to many types of memory loss, including impairment of working, spatial and object recognition memoreis. Curcumin shows strong anti-oxidoinflammatory activities but it terapathics limited by its low solubility in water and cor...

متن کامل

MR artifacts, heat production, and ferromagnetism of Guglielmi detachable coils.

The Guglielmi detachable coil, a platinum microcoil used in the endovascular treatment of intracranial aneurysms, was studied in vitro for its MR imaging artifacts, heat production, and ferromagnetism. In addition, imaging artifacts were studied in vivo in eight patients who had undergone therapeutic placement of these coils. These devices displayed a very low level of MR artifact and no ferrom...

متن کامل

Diffusion-weighted MR imaging after angioplasty or angioplasty plus stenting of arteries supplying the brain.

BACKGROUND AND PURPOSE There has been concern regarding the safety of revascularization procedures of vessels supplying the brain vessels because of the risk of cerebral embolization during the procedure. We have observed a high incidence of hyperintense lesions on diffusion-weighted MR images of the brain after stenting at the carotid bifurcation. The hypothesis of this study is that diffusion...

متن کامل

MR angiography of an aberrant internal carotid artery.

Pulsatile tinnitus with or without a vascular-appearing retrotympanic mass has a number of possible causes (1]. Previous reports (1-4] have stressed the use of CT for evaluation when the masses are vascular appearing. In addition, Remley et al. [1] concluded that MR imaging has only a secondary role in the evaluation of pulsatile tinnitus because it cannot detect vascular variants such as an ab...

متن کامل

Pii: S0967-2109(98)00154-9

Purpose: This study aimed to relate the level of physical force applied to the arterial wall by atraumatic clamps to the degree of endothelial and wall damage. Methods: Sixteen sheep carotid and femoral arteries were each demarcated into four segments 1 cm apart (total 64 segments). Each segment was clamped for 15 min with a standard angled DeBakey vascular clamp. Four levels of force were gene...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 11 2  شماره 

صفحات  -

تاریخ انتشار 1990