Effects of Stimulus Intensity on Low-Frequency Toneburst Cochlear Microphonic Waveforms
نویسنده
چکیده
This study investigates changes in amplitude and delays in low-frequency toneburst cochlear microphonic (CM) waveforms recorded at the ear canal in response to different stimulus intensities. Ten volunteers aged 20-30 were recruited. Low-frequency CM waveforms at 500 Hz in response to a 14-ms toneburst were recorded from an ear canal electrode using electrocochleography techniques. The data was statistically analyzed in order to confirm whether the differences were significant in the effects of stimulus intensity on the amplitudes and delays of the low-frequency CM waveforms. Electromagnetic interference artifacts can jeopardize CM measurements but such artifacts can be avoided. The CM waveforms can be recorded at the ear canal in response to a toneburst which is longer than that used in ABR measurements. The CM waveforms thus recorded are robust, and the amplitude of CM waveforms is intensity-dependent. In contrast, the delay of CM waveforms is intensity-independent, which is different from neural responses as their delay or latency is intensity-dependent. These findings may be useful for development of the application of CM measurement as a supplementary approach to otoacoustic emission (OAE) measurement in the clinic which is severely affected by background acoustic noise. The development of the application in the assessment of low-frequency cochlear function may become possible if a further series of studies can verify the feasibility, but it is not meant to be a substitute for audiometry or OAE measurements. The measurement of detection threshold of CM waveform responses using growth function approach may become possible in the clinic. The intensity-independent nature of CMs with regards to delay measurements may also become an impacting factor for differential diagnoses and for designing new research studies.
منابع مشابه
Effect of modulation of basilar membrane position on the cochlear microphonic.
Certain characteristics of the extracellular cochlear microphonic (CM) recorded by intracochlear electrodes change in a bimodal manner as a function of prior acoustic exposure, intensity of stimulation, or stimulus frequency. In the present study, it is shown that biasing the basilar membrane position toward scala tympani serves to enhance the CM amplitude when the cochlea is unfatigued, when l...
متن کاملObservations on mastoid versus ear canal recorded cochlear microphonic in newborns and adults.
The cochlear microphonic (CM) may play an important role in the diagnosis of auditory neuropathy (AN) in newborns. However, since the CM tends to mirror the waveform of the acoustic stimulus, conscientious recording methodology must be applied to separate true response from artifact. The difficulty in achieving this separation has limited the clinical usefulness of the CM. In an effort to call ...
متن کاملComparison of the effect of quasitrapezoidal and rectangular pulses on bio- electrical activity, calcium spike properties and afterhyperpolarization potentials of Fl cells of Helix aspersa using intracellular recording
While the effect of changes of stimulus waveform (quasitrapezoidal and rectangular current pulses) on nerve activation is clear, but there is no evidence on the effect of quasitrapezoidal pulses on ionic currents of cellular membrane. In the present study, the effect of depolarizing quasi-trapezoidal current pulses, in comparison with that of depolarizing rectangular current pulses, on firing...
متن کاملSome dualistic properties of the cochlear microphonic.
Within a small frequency range just above the characteristic frequency of a differential electrode pair, cochlear microphonic input-output functions are bimodal, exhibiting two maxima. The more sensitive, low-intensity response has a more limited linear operating range, and it is more labile due to acoustic fatigue or hypoxia. After fatigue or hypoxia, the high-intensity response is revealed. T...
متن کاملResponse pattern based on the amplitude of ear canal recorded cochlear microphonic waveforms across acoustic frequencies in normal hearing subjects.
Low-frequency otoacoustic emissions (OAEs) are often concealed by acoustic background noise such as those from a patient's breathing and from the environment during recording in clinics. When using electrocochleaography (ECochG or ECoG), such as cochlear microphonics (CMs), acoustic background noise do not contaminate the recordings. Our objective is to study the response pattern of CM waveform...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 3 شماره
صفحات -