Wear of the Charité® lumbar intervertebral disc replacement investigated using an electro-mechanical spine simulator

نویسندگان

  • Parshia Moghadas
  • Aziza Mahomed
  • Duncan ET Shepherd
  • David WL Hukins
چکیده

The Charité(®) lumbar intervertebral disc replacement was subjected to wear testing in an electro-mechanical spine simulator. Sinusoidally varying compression (0.6-2 kN, frequency 2 Hz), rotation (±2°, frequency 1 Hz), flexion-extension (6° to -3°, frequency 1 Hz) and lateral bending (±2°, frequency 1 Hz) were applied out of phase to specimens immersed in diluted calf serum at 37 °C. The mass of the ultra-high-molecular weight polyethylene component of the device was measured at intervals of 0.5, 1, 2, 3, 4 and 5 million cycles; its volume was also measured by micro-computed tomography. Total mass and volume losses were 60.3 ± 4.6 mg (mean ± standard deviation) and 64.6 ± 6.0 mm(3). Corresponding wear rates were 12.0 ± 1.4 mg per million cycles and 12.8 ± 1.2 mm(3) per million cycles; the rate of loss of volume corresponds to a mass loss of 11.9 ± 1.1 mg per million cycles, that is, the two sets of measurements of wear agree closely. Wear rates also agree closely with measurements made in another laboratory using the same protocol but using a conventional mechanical spine simulator.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The activL® Artificial Disc: a next-generation motion-preserving implant for chronic lumbar discogenic pain

Degeneration of the lumbar intervertebral discs is a leading cause of chronic low back pain in adults. Treatment options for patients with chronic lumbar discogenic pain unresponsive to conservative management include total disc replacement (TDR) or lumbar fusion. Until recently, only two lumbar TDRs had been approved by the US Food and Drug Administration - the Charité Artificial Disc in 2004 ...

متن کامل

Clinical study to evaluate the safety and effectiveness of the Aesculap Activ-L™ artificial disc in the treatment of degenerative disc disease

BACKGROUND The objective of this clinical study is to evaluate the safety and effectiveness of the Activ-L Artificial Disc for treatment of single-level degenerative disc disease of the lumbar spine in patients who have been unresponsive to at least six months of prior conservative care. The hypothesis of the study is that the Activ-L Disc is non-inferior to the control (the Charité Artificial ...

متن کامل

Kineflex lumbar artificial disc versus Charité lumbar total disc replacement for the treatment of degenerative disc disease: A randomized non-inferiority trial with minimum of 2 years’ follow-up

BACKGROUND The Kineflex lumbar artificial disc replacement device (SpinalMotion, Mountain View, California) is a semiconstrained, posterior center of rotation, metal-on-metal intervertebral disc prosthesis. We performed a prospective, randomized, non-inferiority trial comparing the Kineflex Disc with the Food and Drug Administration (FDA)-approved Charité device (DePuy Spine, Raynham, Massachus...

متن کامل

Wear simulation of a polyethylene-on-metal cervical total disc replacement under different concentrations of bovine serum lubricant.

Metal-on-polyethylene total disc replacements have been an alternative to spinal fusion in the lumbar spine under certain indications for more than a decade. Recently, cervical total disc replacement has also become an alternative to cervical fusion. Knowledge acquired from years of in vitro simulator studies on other joint replacements has highlighted the risks associated with premature wear d...

متن کامل

Estimation of spinal loads using a detailed finite element model of the L4-L5 lumbar segment derived by medical imaging kinematics; a feasibility study

Introduction: Low back pain is the most prevalent orthopedic disorder and the first main cause of poor working functionality in developed as wells as many developing countries. In Absence of noninvasive in vivo measurement approaches, biomechanical models are used to estimate mechanical loads on human joints during physical activities. To estimate joint loads via musculoskelet...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 229  شماره 

صفحات  -

تاریخ انتشار 2015